Metrohm 746 VA Trace Analyzer User Manual
Page 382

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-231
Adjustable parameters:
DPMODE
t.step
t.meas
t.pulse
U.ampl
SWEEP
U.start
U.end
U.step
F (Hg)
t.drop
t
U
t
U.step
U.start
I
U
t.meas
(I
1
)
U.peak
I.peak
U.ampl
t.step
t.pulse
t.meas
(I
2
) t.meas
(I
1
) t.meas
(I
2
)
I = I
2
- I
1
Comments
•
The
DPMODE
measurement mode can be used with all types of
electrodes together with the instructions
MEAS
,
SWEEP
,
DSWEEP
,
FSWEEP
or
TSWEEP
.
•
The electrode type
DME
,
SMDE
,
HMDE
or
RDE
must be entered before
the
DPMODE
measurement mode.
•
If the
DPMODE
measurement mode is used with the voltage sweep
SWEEP
and the electrode type
HMDE
, the potentiostat normally oper-
ates in the operating mode "slow". A switch is made to the operating
mode "fast" when the pulse time
t.pulse
< 30.05 ms.
•
If the
DPMODE
measurement mode is used with
SWEEP
and the
electrode type
SMDE
or
DME
, the setting of the potentiostat depends
on the measurement profile. Normally, a switch is also made from
"slow" to "fast" when the pulse time
t.pulse
< 30.05 ms. With very
small signals, however, the operating mode "slow" is used and with
very large signals the operating mode "fast", irrespective of the pulse
time.
•
If the
DPMODE
measurement mode is used with the voltage sweep
SWEEP
and the electrode type
RDE
, the potentiostat always operates in
the operating mode "fast".
•
If the
DPMODE
measurement mode is used with the voltage sweep
FSWEEP
, the potentiostat always operates in the operating mode "fast".
This mode is recommended especially for adsorptive stripping tech-
niques (e.g. Ni and Co with DMG).
•
The quantity
t.drop
important for the definition of
t.step
depends
on the type of electrode selected:
DME
t.drop = 20 ms
HMDE and RDE
t.drop = 0 ms
SMDE
t.drop = Drop.size
×× 40 ms
see also
SWEEP, FSWEEP, DME, HMDE, SMDE, RDE
Area (Hg)