beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 319

background image

5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-168

Parameter

Range; Default value / Meaning

Method

Method name

Title

Method title

Remark 1

Remark 1 regarding method

Remark 2

Remark 2 regarding method

The following reports can be selected:

Report Short

Short report with

Final results

which have

been defined on the "CALCULATIONS" page. This
report is identical with the subpage "Calculations"
(see section 5.5.2).

Report Eval

Evaluation report with detailed results of the peak
or wave evaluation for all substances. This report
is identical with the subpage "Evaluations" (see

section 5.5.2).

Report Full

Full report with the following outputs:

Detailed results of the peak or wave evaluation
for all substances (subpage "Evaluations", see

section 5.5.2)

Results of all substance calibrations (subpage
"Calibrations", see section 5.5.2)

Solutions (dialog page "SOLUTIONS", see

section 5.3.4)

Common variables (dialog page "COMMON
VARIABLES", see section 5.4.5)

Final results (subpage "Calculations", see sec-

tion 5.5.2)

Report MethSpc

Method report with output of all dialog pages of
the group "Method Specifications":

Dialog page "OPERATION SEQUENCE"

Dialog page "SEGMENTS" (all segments)

Dialog page "DOCUMENTATION"

Dialog page "SUBSTANCES" (all substances)

Dialog page "CALCULATION"

Report Mpts

Measured point report with output of the list of
measured points of segments. It is necessary to
enter the name of the substance or segment and
the VR number:

Sbst:

Output of the original measured values of the
segment associated with the specified substance.

Sgmt:

Output of the original measured values of the
specified segment.

VR: xx

Selection of the single measurement whose list of
measured points should be put out by specifica-
tion of the VR number (V = variation, R = replica-
tion). On entry of

*

, all variations or replications

are put out.