Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. DOS>M ..... Soln.name ........ V.add ........mL

Function

Dispense auxiliary solution into measuring vessel (MEAS).

Parameter

Range; Default value / Meaning

Soln.name

8 characters; -

Name of the auxiliary solution

The entry of a solution name is mandatory only when the auxiliary
solution should be added automatically using a 685 Dosimat, a 700
Dosino or the 695 Autosampler. In this case the solution name

Soln.name

must be entered in the parameter of the same name on the

"SOLUTIONS" page and assigned a position

Pos.

(see section 5.3.4).

Without entry of a name for

Soln.name

in the

DOS>M

instruction, without

entry of this name on the "SOLUTIONS" page or without a position
assignment, the auxiliary solution must be dispensed manually or with
the 6.5611.010 Pipetting equipment (the prompt

DOS "Soln.name"

"V.add" > M

then appears during the program run).

V.add

0.010...50.000 mL; 1.000 mL

Solution volume that should be dispensed

Volume of the auxiliary solution that should be dispensed into the
measuring vessel.

Description

The

DOS>M

instruction is used for the manual or automatic addition of

auxiliary solutions (e.g. supporting electrolyte, reagent solution) to the
measuring vessel (MEAS) on the 747 VA Stand.

The automatic addition can be performed with a 685 Dosimat, a 700
Dosino or directly with the 695 Autosampler. The appropriate position of
the auxiliary solution must be specified on the "SOLUTIONS" page (see

Soln.name

) for this.

Comments