3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-46

3.10.5 Installation of Dosinos for rinsing the measuring vessel

It is possible to rinse the measuring vessel at the 747 VA Stand manually via dialog

page "STAND" or with the

RINSE

command by means of two 700 Dosinos with

50 mL Dosing units. To do this, the following accessories have to be ordered in ad-

dition to the standard equipment of the 747 VA Stand (see also section 9.4):

• 2 ×× 2.700.0010 Dosino

• 1 ×× 6.5323.000 Rinsing equipment

The two Dosinos and the accessories have to be installed as follows (see Fig. 22):

1 Install accessories at 747 VA Stand

Install accessories for rinsing and siphoning off at the 747 VA Stand acc.

to section 3.3.2 point 8 (see Fig. 10).

Install 6.1456.210 measuring vessel at the 747 VA Stand.

2 Prepare containers

Unscrew red filling connections from 6.1621.000 Containers.

Using a funnel, add max. 10 L of rinsing solution (normally acidified

ultrapure water with 100

µ

L conc. HCl/L) to the storage container through

its small vent opening.

3 Install Dosinos on containers

Screw a 6.1618.050 Threaded adapter to each container.

Screw a 6.1829.020 Suction tubing (L = 50 cm) to the bottom of the two

6.1567.250 Dosing units (see section 2.1 of the 700 Instructions for use).

Screw the 6.1567.250 Dosing units to the 6.1618.050 Threaded adapters

screwed to the containers.

Mount a 700 Dosino on each 6.1567.250 Dosing unit (procedure, see

section 2.3 of the 700 Instructions for use).

4 Setting up the containers

Set up the waste and storage containers at the desired location next to

the 747 VA Stand (the max. possible distance is limited by the length

1.5 m of the connecting cables of the Dosinos).

5 Define rinsing Dosinos on 746 VA Trace Analyzer

Switch on 746 VA Trace Analyzer using the mains switch 11

11.

Select dialog page "SOLUTIONS".

Add a new solution with

Soln.name

=

RINSE

and

Pos.

=

Do.3

.

Switch off 746 VA Trace Analyzer.

6 Connect rinsing Dosino to 747 VA Stand

Using either a 6.1805.120 FEP Tubing (L = 1 m) or two 6.1805.120 FEP

Tubings and a 6.1808.000 Connecting sleeve, connect connection 1 of

the 6.1567.250 Dosing unit of the rinsing Dosino with the 6.1808.000

Coupling connected to the FEP tubing 90

90 and mounted on the 747 VA

Stand (see Fig. 10).

Attach rinsing Dosino using the cable connected to the instrument to

socket 19

19 (Dos. 3) of the 746 VA Trace Analyzer (see Fig. 2 or Fig. 4).