7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-26
Error message
Meaning/Causes
Rectification
Exit
STORAGE - ERROR 4:
Files are identical
Copying the file with the
specified name is not possible
as a file with the same name
already exists
Change file name
E/C
STORAGE - ERROR 5:
Invalid destination for
multiple files
Inadmissible destination for the
copying of several files
Copy one file only
E/C
STORAGE - ERROR 6:
Invalid destinations
for methods
Inadmissible destination for the
copying of methods
Change destination for copying
E/C
STORAGE - ERROR 7:
Invalid destinations
for determinations
Inadmissible destination for the
copying of determinations
Change destination for copying
E/C
STORAGE - ERROR 8:
Invalid determination
Inadmissible destination for the
copying of files
Change destination for copying
E/C
STORAGE - ERROR 9:
RS232 Receive mode off
Copying from RS232 interface
not possible as receive mode
disabled
Enable receive mode on the ”VA TRACE
ANALYZER” page (
Receive mode =
on
) and restart copying
E/C
STORAGE - ERROR 10:
RS232 Keyboard send
mode still active
Copying to or from RS interface
not possible as
Remote
keyboard = on
is set
Set
Remote keyboard = off
on the
”VA TRACE ANALYZER” page and
restart copying
E/C
STORAGE - ERROR 11:
RS232 RS232 interface
still active
Copying from RS interface not
possible as it is still active
−
Wait a short time then press
(retry) and try again
−
(abort) and reset RS interface with the
softkey
on the ”VA
TRACE ANALYZER” page
r/a
STORAGE - ERROR 12:
Invalid file name
The inputted file name contains
inadmissible characters or a
wrong extension
Correct file name
E/C
STORAGE - ERROR 13:
RS232 interface,
invalid internal
parameters
Invalid internal parameters on
copying to or from RS interface
−
Wait a short time then press
(retry) and try again
−
(abort) and reset RS interface with the
softkey
on the ”VA
TRACE ANALYZER” page
E/C
STORAGE - ERROR 14:
Copy terminated with
error
Copying ended with error
Check which files have not been
copied, try again
E/C
STORAGE - ERROR 15:
Unexpected file
blocksize
Unexpected block size in
copying from RS interface
−
Wait a short time then press
(retry)
−
(abort) and try copying again
r/a
STORAGE - ERROR 16:
Memory blocksize
violation
Error in the data expansion in
copying from RS interface
−
Wait a short time then press
(retry)
−
(abort) and try copying again
r/a
STORAGE - ERROR 17:
Undefined error
Non-specified storage error
Press Reset key
E/C
SUBSTANCES - ERROR 1:
Range min greater than
Range max
The lower limit of the working
range is greater than the upper
limit
Enter
Range min
and
Range max
parameters on the ”SUBSTANCES”
page properly
Esc
SUBSTANCES - ERROR 2:
Param. Curve type
inconsistent with
Technique
The specified entry for the
Curve type
parameter is not
admissible for the selected
calibration technique
Change
Curve type
parameter on the
”SUBSTANCES” page
Esc