7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-26

Error message

Meaning/Causes

Rectification

Exit

STORAGE - ERROR 4:
Files are identical

Copying the file with the

specified name is not possible

as a file with the same name

already exists

Change file name

E/C

STORAGE - ERROR 5:
Invalid destination for
multiple files

Inadmissible destination for the

copying of several files

Copy one file only

E/C

STORAGE - ERROR 6:
Invalid destinations
for methods

Inadmissible destination for the

copying of methods

Change destination for copying

E/C

STORAGE - ERROR 7:
Invalid destinations
for determinations

Inadmissible destination for the

copying of determinations

Change destination for copying

E/C

STORAGE - ERROR 8:
Invalid determination

Inadmissible destination for the

copying of files

Change destination for copying

E/C

STORAGE - ERROR 9:
RS232 Receive mode off

Copying from RS232 interface

not possible as receive mode

disabled

Enable receive mode on the ”VA TRACE

ANALYZER” page (

Receive mode =

on

) and restart copying

E/C

STORAGE - ERROR 10:
RS232 Keyboard send
mode still active

Copying to or from RS interface

not possible as

Remote

keyboard = on

is set

Set

Remote keyboard = off

on the

”VA TRACE ANALYZER” page and

restart copying

E/C

STORAGE - ERROR 11:
RS232 RS232 interface
still active

Copying from RS interface not

possible as it is still active

Wait a short time then press

(retry) and try again

If the error reappears, press

(abort) and reset RS interface with the

softkey

on the ”VA

TRACE ANALYZER” page

r/a

STORAGE - ERROR 12:
Invalid file name

The inputted file name contains

inadmissible characters or a

wrong extension

Correct file name

E/C

STORAGE - ERROR 13:
RS232 interface,
invalid internal
parameters

Invalid internal parameters on

copying to or from RS interface

Wait a short time then press

(retry) and try again

If the error reappears, press

(abort) and reset RS interface with the

softkey

on the ”VA

TRACE ANALYZER” page

E/C

STORAGE - ERROR 14:
Copy terminated with
error

Copying ended with error

Check which files have not been

copied, try again

E/C

STORAGE - ERROR 15:
Unexpected file
blocksize

Unexpected block size in

copying from RS interface

Wait a short time then press

(retry)

If the error reappears, press

(abort) and try copying again

r/a

STORAGE - ERROR 16:
Memory blocksize
violation

Error in the data expansion in

copying from RS interface

Wait a short time then press

(retry)

If the error reappears, press

(abort) and try copying again

r/a

STORAGE - ERROR 17:
Undefined error

Non-specified storage error

Press Reset key

E/C

SUBSTANCES - ERROR 1:
Range min greater than
Range max

The lower limit of the working

range is greater than the upper

limit

Enter

Range min

and

Range max

parameters on the ”SUBSTANCES”

page properly

Esc

SUBSTANCES - ERROR 2:
Param. Curve type
inconsistent with
Technique

The specified entry for the

Curve type

parameter is not

admissible for the selected

calibration technique

Change

Curve type

parameter on the

”SUBSTANCES” page

Esc