Metrohm 746 VA Trace Analyzer User Manual
Page 535

7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-36
Peak displacement
•
Check and adjust the pH of the solution.
•
Check electrolyte composition and correct if necessary. Use a buffer solution
instead of an acid.
•
Carry out a standard addition to check whether the correct peak has been
evaluated.
•
Organic components interfere with the analysis: carry out a UV digestion or
other suitable sample preparation.
• Enter a new half-wave potential in the instrument and recalculate the results.
•
Check reference electrode (see section 3.6).
No peak found
•
The peak is only displaced: adjust the half-wave potential and recalculate the
results.
•
The sample concentration is too low: increase the sample volume or the
amount of sample.
•
The concentration of the ion to be determined is too low: use HMDE (inverse
voltammetry) instead of DME or SMDE.
•
Are the initial and final potentials correct?
Peak is in the highest
µ
A range
•
The concentration of the ion to be determined is too high: reduce the sample
volume and carry out the analysis again.
•
The preconcentration time under
MEAS
is too high, reduce it and try the
analysis again.
Double peak
•
Check MME. If necessary, change capillary (see section 3.4.8) or replace
sealing needle (see section 3.4.9).
•
Organic components interfere with the analysis: carry out a UV digestion or
other suitable sample preparation.
•
If a second element is present at the same potential: add this element to the
sample and carry out the analysis again. If the second peak has become
higher then the second element is present.
•
Has any substance been precipitated out in the measuring vessel (e.g. lead
perchlorate standard with KCl as electrolyte)?
•
Try out eluents with different compositions (addition of complex formers).
•
Check analysis parameters.
•
Try another measurement mode like AC1. If one substance is reversible and
the second one irreversible, only the reversible substance is detected by
AC1.
Standard addition peaks displaced
•
Standard solutions have been made too acidic.
•
Buffering capacity of the electrolyte is not sufficient: increase electrolyte vol-
ume.