5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-290

SMPL/M

Page:

O, S

Stand:

M

Abbreviation:

SMPL/M

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SMPL/M ..... V.fraction ........mL V.total .........L

Function

Sample volume has been added to the measuring vessel (MEAS) on the
747 VA Stand.

Parameter

Range; Default value / Meaning

V.fraction

0.010...50.000 mL; -

Fractional volume

With diluted samples, here the volume fraction of the total volume

V.total

which has been added to the measuring vessel must be

entered. The following condition holds:

V.fraction

≤≤ V.total

V.total

0.10 mL...100000 L; -

Total volume

With diluted samples, here the total volume of sample and dilution
solution must be entered. The following condition holds:

V.total

≥≥ V.fraction

Description

The

SMPL/M

instruction defines the sample volume in the measuring

vessel (MEAS) on the 747 VA Stand.

With undiluted samples (

V.fraction

and

V.total

blank), the sample

volume is given either by

S0

(on entry of

S0

as a volume) or by

S0/S2

(on entry of

S0

as a weight,

S2

is the density in g/mL).

With diluted samples (

V.fraction > 0

and

V.total

≥≥ V.fraction

), the

sample volume is given by

V.fraction

, irrespective of the entry of

S0

.

You will find further details on the sample size in the description of the
parameter

Sample size/S0

in section 5.4.4.

Comments

The

SMPL/M

instruction can not be used in operation with the 695

Autosampler.

see also

SMPL>M, SMPL/F