225 dme, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual

Page 376

background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-225

DME

Page:

O, S

Stand:

M, A

Abbreviation:

DM

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. DME .....

Function

Switch on Dropping Mercury Electrode (DME).

Description

The

DME

instruction is used to switch on the 6.1246.020 Multi-Mode

Electrode as a dropping mercury electrode, the mercury starts to flow out
freely. In a subsequent voltage sweep, the Hg drops are knocked off by
a tapping mechanism after the time

t.step

set in the measurement

mode.

t.step

t

F (Hg)

The set drop time

t.drop

important for certain parameter settings is

20 ms for the DME.

Comments

Within a program, the DME is not switched off until the

ØMEAS

instruction or the selection of a different electrode type (

HMDE

,

SMDE

or

RDE

). The DME is also switched off by

END

in the main program on

the "OP.SEQUENCE" page, but not by

END

in the segment.

The DME can be used for the following measurement modes and
sweeps:

DCTMODE

with

SWEEP

and

SSWEEP

,

DPMODE

with

SWEEP

and

FSWEEP

,

SQWMODE

with

SWEEP

and

FSWEEP

,

AC1MODE

with

SWEEP

,

AC2MODE

with

SWEEP

.

The following are not compatible with the DME:

CYCMODE

with

(CSWEEP

,

PSAMODE

with

PSWEEP

as well as

TSWEEP

.

An advantage of the DME compared with the SMDE is that the MME
capillary is subjected to less mechanical stress.

A disadvantage of the DME compared with the SMDE and HMDE is
the higher mercury consumption and the lower sensitivity as the
electrode surface and hence the baseline constantly changes during
the measurement phase.

see also

HMDE, SMDE, RDE

Area (Hg)