Metrohm 746 VA Trace Analyzer User Manual

Page 287

background image

5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-136

Parameter

Range; Default value / Meaning

Baseline

display only

Baseline

Title for those parameters which concern calculation of the baseline.
These parameters are not displayed for substances recorded with
measurement mode PSAMODE.

Type

linear, polynom, expon., AC2 rel., AC2 abs.; linear

Baseline type

The following possibilities are available for approximation of the
baseline and the evaluation of peaks/waves (see also section 6.7):

linear

Linear baseline with DPMODE, SQWMODE, AC1MODE,

DCTMODE (with SSWEEP):
The

linear

type baseline is determined using the tangent

method. After definition of the peak base points

dU.front

and

dU.rear

(automatically or by manual entry), starting

from the base points an iterative search within a fixed
range (inward: up to 1 point before the peak maximum;
outward: ca. 20 mV) is made for a common slope. The
two points of tangency are connected to produce the lin-
ear baseline.
If no common slope is found, the linear baseline is drawn
directly through the two base points.

linear, whole

dU.front dU.rear

Linear baseline with DCTMODE (with SWEEP):
Wave-shaped curves recorded with DCTMODE are al-
ways evaluated with the tangent method

Type = linear

.

The base points

dU.front

and

dU.rear

and the associ-

ated slopes

S.front

and

S.rear

can be determined

automatically or entered manually. In contrast to peaks
with a linear baseline, the slopes may be different.

S.front

S.rear

U.wave

I.wave

dU.front dU.rear

linear, whole (DCT)

Point of tangency

(rear)

Point of tangency

(front)

Base point

(front)

Range for

tangent search

Range for

tangent search

Base point

(rear)

Base point

(rear)

Base point

(front)