6.9 Content calculation

746 VA Trace Analyzer / 747 VA Stand

6-27

Working in the nonlinear range

If the substance content is in the nonlinear range, several standard additions

have to be performed to determine the nonlinear curve shape exactly.

Standard addition ratio 1:2 to 1:5

The optimum standard addition ratio for the entire standard addition is from

1:2 to 1:5, i.e. the sum of all standard addition amounts should be 2 to 5

times the amount of sample in the measuring vessel. This can be easily

checked later with the parameters

Mass

and

Add.mass

put out on the dialog

page "RESULTS/Evaluations".

Consideration of blank values

Any blank values must be determined separately and subtracted in the for-

mula calculation on the dialog page "CALCULATIONS".

6.9.5

Calibration curve

The content determination using a calibration curve is performed in two steps:

– First the relation between the mass concentration

ρ

eff

eff

of a sub-

stance and the evaluation quantity

EV

is determined by measur-

ing different reference solutions.

– Finally, the sample is measured and its mass concentration

ρ

eff,s

eff,s

determined using the recorded calibration curve.

The recording of a calibration curve is programmed with the following instructions

and parameters on the "OP.SEQUENCE" page (see section 5.7.3):

(CAL...CAL)n

Loop for recording the calibration curve

n

Number of calibration solutions – 1

CSOLN...

Addition of the calibration solutions

The remaining parameters needed for the calibration curve are entered on the dia-

log page "SUBSTANCES/Calibration" (see section 5.6.3). The most important of

these are:

Technique

Selection of the calibration technique

cal.crv.

Calibration curve

Curve type

Curve type

linear

Straight line of type

y = a + bx

0/linear

Straight line of type

y = bx

nonlin

Curve of type

y = a + bx + cx

4

0/nonlin

Curve of type

y = bx + cx

4

V0.CSoln.

Addition volume of the calibration solutions

CSoln.name

Name of the calibration solutions

Mass conc.

Mass concentration of the calibration solutions

In the content determination using a calibration curve, the following procedure is

used to calculate the sought mass concentration

ρ

eff,s

eff,s

of the sample (see also ex-

ample in section 4.3):