beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 542

background image

7.5 Vagaries of the chemistry

746 VA Trace Analyzer / 747 VA Stand

7-43

7.5.3

Unsuitable bridging electrolyte solution in the reference
electrode

When choosing the bridging electrolyte in the reference electrode, possible compli-

cations with the substances present in the analysis solution must be taken into ac-

count.
With regard to the bridging electrolyte solution KCl 3 mol/L used in many cases,

the following are examples of disturbances which can appear:

• Precipitation of KClO

4

in the ceramic diaphragm

with supporting electrolytes containing HClO

4

With partial blockage, inexplicable side peaks can appear, with complete

blockage one of the following error messages appears:

VA STAND - ERROR 2: Working electrode faulty
VA STAND - ERROR 4: Reference electrode faulty

To avoid such precipitations, with analysis solutions containing HClO

4

a

bridging electrolyte solution free from alkalis or alkaline earths (e.g. acetate

buffer) must be used.

Introduction of Cl

through KCl outflow from

the reference electrode

The outflow of bridging electrolyte from the ceramic diaphragm of the

6.1245.010 Electrolyte vessel (part of the reference electrode) is 2...5

µ

L/h.

The Cl

flowing into the analysis solution can interfere with the determination

of vitamin C or the determination of Cu (see also section 7.5.6).
The use of Cl

free bridging electrolyte solutions (e.g. KNO

3

sat.) is rec-

ommended as a countermeasure.

7.5.4

Overcharging of the working electrode

Under unfavorable conditions (high concentrations and/or long enrichment times),

the enrichment of species at polarized electrodes leads to overcharging phenom-

ena such as non-linear standard addition curves or splitting into multiple peaks

which are caused by saturation and different deposition forms.
A shorter enrichment time usually solves the problem. The following rule of

thumb holds: In general, enrichment should not be carried out except in solutions

with a mass concentration

ρ < 0.5 mg/L (= 0.5 ppm). In several cases work can be

carried out without enrichment even with concentrations

ρ > 100

µ

g/L (e.g. DP

voltammetry at the HMDE or also at the DME).
The effects of an enrichment time which is too long are shown by the following two

examples:

• Nickel and cobalt determination in the trace region by cathodic

adsorption stripping voltammetry (with dimethylglyoxime com-

plexes)

Prolongation of the enrichment time from 30 s to 120 s (keeping all other

measurement parameters constant) leads to non-linear standard additions

and in the case of nickel also to shifts in the peak maximum: