3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-42

3.10.2 Tubing connection for addition of standard solutions

For the addition of standard solutions in the

µ

L-range into the measuring vessel of

the 747 VA Stand the 4-way microtip 47

47 (6.1824.000) should be used. It is fitted

with 4 lengths of PTFE tubing with connection nipples for direct attachment to the

Dosing unit of the 700 Dosino. To ready the 700 Dosino for automatic dispensing,

proceed as follows:

1 Mount Dosing unit on 700 Dosino

Procedure, see section 2.2 of 700 Instructions for Use.

2 Insert 4-way microtip in 747 VA Stand (see section 3.3.2)

Remove stopper from nipple 44

44 and insert 4-way microtip 47

47 into nipple

44

44 as far as it will go (see Figs 7 and 9).

Tighten nipple 44

44 using a 6.2739.000 Wrench (accessory of 6.3030.XXX

Dosing unit) so that the 4-way microtip 47

47 can no longer move.

Pull the 4 lengths of PTFE tubing of the 4-way microtip 47

47 in succession

from above through the opening 86

86.

3 Connect PTFE tubing to Dosing unit

Screw connection nipple of the PTFE tubing of the 4-way microtip 47

47 onto

connection 1 of the Dosing Unit mounted on the Dosino.

To lengthen the tubing connection the 6.1805.100 FEP tubing (L = 40

cm, accessory of Dosing Unit) can be inserted between PTFE tubing and

connection 1 of the Dosing Unit using the 1808.000 coupling.

4 Close unused PTFE tubings

Screw a 6.1808.000 Coupling (accessory of 747 VA Stand) on each

unused PTFE tubing of the 4-way microtip 47

47.

Screw a 6.1446.040 Dummy stopper (accessory of 747 VA Stand) on

each 6.1808.000 Coupling.

5 Initialize Dosino(s)

Switch on 746 VA Trace Analyzer using the mains switch 11

11.

Select dialog page "DOSIMATS".

If volumes

50

µ

L should be dispensed, set parameter

V.init

to

0.05 mL

.

Set parameter

V.tube in

to

0.7 mL

.

Set parameter

V.tube out

to

0.2 mL

.

Press softkey

for initializing all attached Dosinos or select

desired Dosino with the cursor keys and press softkey

:

The cylinder is filled and excess solution is expelled to the measuring

vessel. With standard accessories, all tubings to the measuring vessel will

have been filled at the end of the initializing procedure.