Metrohm 746 VA Trace Analyzer User Manual

Page 568

background image

8.2 747 VA Stand

746 VA Trace Analyzer / 747 VA Stand

8-7

Measuring amplifier

Max. phase error

1° at 250 Hz

Resolution

24 pA

Multi-mode electrode MME (working electrode WE)

Designation

6.1246.020

Electrode types

DME (dropping mercury electrode)
HMDE (hanging mercury drop electrode)
SMDE (static mercury drop electrode)

Drop surface

0.15 ... 0.60 mm

2

(DME and SMDE)

Glass capillary

6.1226.030 (set of 10)
internal diameter = 0.05 mm

Mercury reservoir

6 mL

81.2 g

sufficient for ca. 200'000 Hg drops

Auxiliary power

inert gas (generally nitrogen N

2

)

p = 1

±

0.2 bar

Rotating disk electrode RDE (working electrode WE, option)

Construction

6.1246.000 Drive shaft + screw-on

6.1204.XXX Electrode tip

Electrode tips

6.1204.100 Ultra trace graphite

6.1204.110 glassy carbon (GC)

6.1204.120 platinum (Pt)

6.1204.130 silver (Ag)

6.1204.140 gold (Au)

Disk diameter

2.0 +0 / –0.05 mm

Radial eccentricity

0.2 mm

Regeneration

with 6.2802.000 Polishing kit

Rotational speed

200, 400, 600, ... , 3000 min

–1

Speed constancy

±

2 %

Reference electrode (RE)

Construction

double-junction;

6.0728.0X0 Ag/AgCl Reference system +

6.1245.010 Electrolyte vessel to be filled by user

Reference system

Ag/AgCl/c(KCl) = 3 mol/L

Diaphragm

ceramic diaphragm

diameter = 3 mm

Auxiliary electrode (AE)

Pt auxiliary electrode

6.0343.000 Platinum electrode

GC auxiliary electrode

(option)

6.1241.020 Electrode holder + 6.1247.000 Glassy

carbon tip