Metrohm 746 VA Trace Analyzer User Manual
Page 568

8.2 747 VA Stand
746 VA Trace Analyzer / 747 VA Stand
8-7
Measuring amplifier
Max. phase error
1° at 250 Hz
Resolution
≈
24 pA
Multi-mode electrode MME (working electrode WE)
Designation
6.1246.020
Electrode types
DME (dropping mercury electrode)
HMDE (hanging mercury drop electrode)
SMDE (static mercury drop electrode)
Drop surface
0.15 ... 0.60 mm
2
(DME and SMDE)
Glass capillary
6.1226.030 (set of 10)
internal diameter = 0.05 mm
Mercury reservoir
6 mL
≅
81.2 g
sufficient for ca. 200'000 Hg drops
Auxiliary power
inert gas (generally nitrogen N
2
)
p = 1
±
0.2 bar
Rotating disk electrode RDE (working electrode WE, option)
Construction
6.1246.000 Drive shaft + screw-on
6.1204.XXX Electrode tip
Electrode tips
6.1204.100 Ultra trace graphite
6.1204.110 glassy carbon (GC)
6.1204.120 platinum (Pt)
6.1204.130 silver (Ag)
6.1204.140 gold (Au)
Disk diameter
2.0 +0 / –0.05 mm
Radial eccentricity
≤
0.2 mm
Regeneration
with 6.2802.000 Polishing kit
Rotational speed
200, 400, 600, ... , 3000 min
–1
Speed constancy
±
2 %
Reference electrode (RE)
Construction
double-junction;
6.0728.0X0 Ag/AgCl Reference system +
6.1245.010 Electrolyte vessel to be filled by user
Reference system
Ag/AgCl/c(KCl) = 3 mol/L
Diaphragm
ceramic diaphragm
diameter = 3 mm
Auxiliary electrode (AE)
Pt auxiliary electrode
6.0343.000 Platinum electrode
GC auxiliary electrode
(option)
6.1241.020 Electrode holder + 6.1247.000 Glassy
carbon tip