5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-283

SETCTRL

Page:

O, S

Stand:

M, A

Abbreviation:

SET

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SETCTRL ..... Code ........ Message ................

Function

Set the output control lines 1...8.

Parameter

Range; Default value / Meaning

Code

1, 0, *; *

Code for output control lines

8-digit numeric code for setting the output control lines 1...8. The
following values can be entered for each line:

0

off, inactive (open)

1

on, active

(0 V)

*

leave status unchanged

Message

16 characters;

Message for output control lines

Possibility for entry of a remark concerning the command.
If this field is edited, 4 preset messages for operation of the 708 Sam-
pling Unit are displayed on the softkeys. If such a message is entered
by pressing one of the softkeys, the last three digits of the code are
simultaneously changed:

FILL PUMP ON

Ю

Ю Code = *****101

Injection valve to "FILL",
pump on

FILL PUMP OFF

Ю

Ю Code = *****100

Injection valve to "FILL",
pump off

INJECT PUMP ON

Ю

Ю Code = *****011

Injection valve to "INJECT",
pump on

INJECT PUMP OFF Ю

Ю Code = *****010

Injection valve to "INJECT",
pump off

Description

The

SETCTRL

instruction is used to set the output control lines 1...8 of the

remote interface "Control Lines" to the status preset in

Code

. Functions

can thus be initiated at external devices attached to the remote interface.

Comments

The attachment of devices to the remote interface is described in section

3.13.

Example

see

ADD/M

see also

SCANCTRL