283 setctrl, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 434

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-283
SETCTRL
Page:
O, S
Stand:
M, A
Abbreviation:
SET
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SETCTRL ..... Code ........ Message ................
Function
Set the output control lines 1...8.
Parameter
Range; Default value / Meaning
Code
1, 0, *; *
Code for output control lines
8-digit numeric code for setting the output control lines 1...8. The
following values can be entered for each line:
0
off, inactive (open)
1
on, active
(0 V)
*
leave status unchanged
Message
16 characters;
Message for output control lines
Possibility for entry of a remark concerning the command.
If this field is edited, 4 preset messages for operation of the 708 Sam-
pling Unit are displayed on the softkeys. If such a message is entered
by pressing one of the softkeys, the last three digits of the code are
simultaneously changed:
FILL PUMP ON
Ю
Ю Code = *****101
Injection valve to "FILL",
pump on
FILL PUMP OFF
Ю
Ю Code = *****100
Injection valve to "FILL",
pump off
INJECT PUMP ON
Ю
Ю Code = *****011
Injection valve to "INJECT",
pump on
INJECT PUMP OFF Ю
Ю Code = *****010
Injection valve to "INJECT",
pump off
Description
The
SETCTRL
instruction is used to set the output control lines 1...8 of the
remote interface "Control Lines" to the status preset in
Code
. Functions
can thus be initiated at external devices attached to the remote interface.
Comments
The attachment of devices to the remote interface is described in section
3.13.
Example
see
ADD/M
see also
SCANCTRL