5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-294
The phase dependent differences in the current measurements give
peak-shaped curves which can be evaluated using the baselines of type
linear
,
polynom
or
expon.
(see section 5.6.3).
I
U
U.peak
I.peak
Comments
•
The
SQWMODE
measurement mode can be used with all types of
electrodes together with the instructions
MEAS
,
SWEEP
,
DSWEEP
,
FSWEEP
or
TSWEEP
.
•
The electrode type
DME
,
SMDE
,
HMDE
or
RDE
must be entered before
the
SQWMODE
measurement mode.
•
If the
SQWMODE
measurement mode is used with the voltage sweep
SWEEP
and the electrode type
HMDE
, the potentiostat normally oper-
ates in the operating mode "slow". A switch is made to the operating
mode "fast" when the modulation frequency
Modul.freq.
>120 Hz.
•
If the
SQWMODE
measurement mode is used with the voltage sweep
SWEEP
and the electrode type
SMDE
or
DME
, the setting of the poten-
tiostat depends on the measurement profile. Normally, a switch is
also made from "slow" to "fast" when the modulation frequency
Modul.freq.
>120 Hz. With very small signals, however, the operat-
ing mode "slow" is used and with very large signals the operating
mode "fast", irrespective of the pulse time.
•
If the
SQWMODE
measurement mode is used with the voltage sweep
SWEEP
and the electrode type
RDE
, the potentiostat always operates in
the operating mode "fast".
•
If the
SQWMODE
measurement mode is used with the voltage sweep
FSWEEP
, the potentiostat always operates in the operating mode "fast"
with all electrode types.
•
The quantity
t.drop
important for the definition of
t.step
depends
on the type of electrode selected:
DME
t.drop = 20 ms
HMDE and RDE
t.drop = 0 ms
SMDE
t.drop = Drop.size
×× 40 ms
see also
SWEEP, FSWEEP, DME, SMDE, HMDE, RDE