5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-294

The phase dependent differences in the current measurements give
peak-shaped curves which can be evaluated using the baselines of type

linear

,

polynom

or

expon.

(see section 5.6.3).

I

U

U.peak

I.peak

Comments

The

SQWMODE

measurement mode can be used with all types of

electrodes together with the instructions

MEAS

,

SWEEP

,

DSWEEP

,

FSWEEP

or

TSWEEP

.

The electrode type

DME

,

SMDE

,

HMDE

or

RDE

must be entered before

the

SQWMODE

measurement mode.

If the

SQWMODE

measurement mode is used with the voltage sweep

SWEEP

and the electrode type

HMDE

, the potentiostat normally oper-

ates in the operating mode "slow". A switch is made to the operating
mode "fast" when the modulation frequency

Modul.freq.

>120 Hz.

If the

SQWMODE

measurement mode is used with the voltage sweep

SWEEP

and the electrode type

SMDE

or

DME

, the setting of the poten-

tiostat depends on the measurement profile. Normally, a switch is
also made from "slow" to "fast" when the modulation frequency

Modul.freq.

>120 Hz. With very small signals, however, the operat-

ing mode "slow" is used and with very large signals the operating
mode "fast", irrespective of the pulse time.

If the

SQWMODE

measurement mode is used with the voltage sweep

SWEEP

and the electrode type

RDE

, the potentiostat always operates in

the operating mode "fast".

If the

SQWMODE

measurement mode is used with the voltage sweep

FSWEEP

, the potentiostat always operates in the operating mode "fast"

with all electrode types.

The quantity

t.drop

important for the definition of

t.step

depends

on the type of electrode selected:

DME

t.drop = 20 ms

HMDE and RDE

t.drop = 0 ms

SMDE

t.drop = Drop.size

×× 40 ms

see also

SWEEP, FSWEEP, DME, SMDE, HMDE, RDE

n = Meas.cycles

Σ

(I

n1

– I

n2

)

n

I

=