5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-192

ADD/M

Page:

O, S

Stand:

M

Abbreviation:

ADD/

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. ADD/M ..... Soln.name ........ V.add ........mL

Function

Standard addition solution has been added to the measuring vessel
(MEAS) (e.g. with a 665 Dosimat).

Parameter

Range; Default value / Meaning

Soln.name

8 characters;

Name of the standard addition solution

The name of the standard addition solution must be identical to the
name entered on the "SUBSTANCES/Calibration" page under

Soln.name

(see section 5.6.3). The mass concentration of the standard

addition solution

Mass conc.

must also be entered on this page.

V.add

0.010...50.000 mL; 0.100 mL

Solution volume that has been added

Volume of the standard addition solution that has been added to the
measuring vessel.

Description

The

ADD/M

instruction is needed when a standard addition solution

should be added automatically using a dosing unit (e.g. 665 Dosimat or
725 Dosimat) which, in contrast to the 685 Dosimat or 700 Dosino, can
not be directly controlled by the 746 VA Trace Analyzer. In this case, the
addition must be initiated via the RS232 interface (using the

SEND>RS#

instructions) or via the remote interface (using the

SETCTRL

instructions)

and then confirmed with the

ADD/M

instruction.

Comments

The

ADD/M

instruction must be located within an ADD loop.

In the last execution of the ADD loop, the

ADD/M

instruction is no

longer executed.

The

ADD/M

instruction can not be used in operation with the 695

Autosampler.