5.5.1 MONITORING

746 VA Trace Analyzer / 747 VA Stand

5-77

Parameter

Range; Default value / Meaning

Start interval

0.0...99.9 h; 0.0 h

Time interval for automatic determination start

This parameter enables the automatic start of determinations at
regular time intervals for automatic series determinations with

Auto.batch = on

(e.g. for online analysis).

The time interval set must be > the time needed for the whole deter-
mination, otherwise the next determination is started immediately.

Det.storage

on, off; off

Automatic storage of determinations

The automatic storage of determinations in the determination memory
"DETERMINATIONS" can be switched on or off here.

on

The determinations are automatically transferred to the
determination memory. If the memory is full, the pro-
cedure depends on the setting of the parameter

Full

condition

whereby either the oldest determinations

are deleted (

Full condition = push out

) or the de-

termination series is stopped (

Full condition = stop

,

see section 5.4.2).

off

The determination data are not stored.

Determination

xxxx-xx-xx
xx:xx:xx

display only

Loaded determination

Name of the determination currently loaded in the working memory or
currently running. A name is given to each determination when it is
recorded and comprises the current time on method start and the
extension

.dtm

:

MMDDhhmm.dtm

MM

= month

hh

= hour

DD

= day

mm

= minutes

The first, 8-digit part of the name can be modified on these dialog
pages after copying to the determination memory or the data card,
but the extension

.dtm

must be retained as it characterizes the

determination files.

Below the determination name, the exact time when the determination
was started is displayed.

Modified

display only

Subsequent modification

Date and time when the determination loaded in the working memory
was subsequently modified for the first time (later modifications are
not displayed). The modification date is set only in the case of modifi-
cations on the following dialog pages:

COMMON VARIABLES

Changes to

Workg.com.var

and the

associated

Remarks

OP.SEQUENCE

All changes

SEGMENTS

All changes

MONITORING

Modification of the parameters in the 1

st

sample data line, recalculation with soft-
key

).