285 simpulse, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 436

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-285
SIMPULSE
Page:
O, S
Stand:
M, A
Abbreviation:
SI
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SIMPULSE .....
Function
Simulation of standard additions with the 3.496.8380 VA Simulator.
Description
The
SIMPULSE
instruction is used to put out a total of 100 pulses at the
3.496.8380 VA Simulator which cause an increase in the current
measurement and hence a simulation of standard additions. The VA
Simulator can be reset to the original current level with the
SIMRESET
instruction.
Comments
•
The 3.496.8380 VA Simulator and the associated 3.496.5190
Connecting cable are available as an option. The VA Simulator gen-
erates 2 polarographic current waves with half-wave potentials of ca.
–540 mV and ca. –1070 mV in a defined voltage range. You will find
further details in the 3.496.8380-SB Circuit Description supplied with
the VA Simulator.
•
To obtain a larger current increase, the
SIMPULSE
instruction can also
be entered twice in succession.
Example
Test program for 3 simulated standard additions with the 3.496.8380 VA
Simulator:
Method SIMI .mth OPERATION SEQUENCE
Title Test program
================================================================================
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
1 SMPL/M V.fraction mL V.total L
2 SIMRESET 1.0
3 (ADD
4 SEGMENT Segm.name Seg1
5 SIMPULSE
6 ADD>M Soln.name V.add 0.100 mL
7 ADD)3
8 END
Method SIMI SEGMENT
Seg1
================================================================================
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
1 HMDE Drop size 4 Meas.cell normal
2 DPMODE U.ampl 50 mV t.meas 20.0 ms
t.step 0.20 s t.pulse 40.0 ms
3 SWEEP 50.4 U.start -100 mV U.step 6 mV
U.end -1600 mV Sweep rate 30 mV/s
4 END
see also
SIMRESET