7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-16
Error message
Meaning/Causes
Rectification
Exit
MALLOC - ERROR 1:
System heap, not enough
memory
Insufficient system storage
space available (typical case:
sending large or many files)
This error message is acknowledged
automatically after 10 s waiting time.
r/a
MALLOC - ERROR 2:
Determination heap, not
enough memory
Insufficient storage space
available for determinations
−
Press (abort)
−
Create storage space by deleting old
determinations
−
Restart copying
r/a
MALLOC - ERROR 3:
Method heap, not enough
memory
Insufficient storage space
available for methods
−
Press (abort)
−
Create storage space by deleting old
methods
−
Restart copying
r/a
MALLOC - ERROR 4:
Storage heap, not
enough memory
Insufficient storage space
available for methods or
determinations
−
Press (abort)
−
Create storage space by deleting old
methods or determinations
−
Restart copying
r/a
MALLOC - ERROR 5:
Can not realloc, not
enough memory
Insufficient storage space
available
−
Press (abort)
−
Create storage space by deleting old
methods or determinations
−
Restart copying
r/a
MALLOC - ERROR 6:
Last powerdown failed,
malloc data incon-
sistent
Inadmissible power failure
during data storage (this error
message appears when the
instrument is switched on)
−
Press
−
Store data from method and determi-
nation memory on data card
−
Reinitialize method and data memory
r/a
MALLOC - ERROR 7:
System heap out of
memory
Insufficient system storage
space available
−
Wait a short time, then press
(retry)
−
If error reappears, press (abort);
an instrument reset is initiated auto-
matically
−
Note:
If this error message appears, the
pressed (danger of crash)
r/a
MEASURE - ERROR 1:
Missing method
The method called with
Method.call
is missing in the
method memory
Load method in method memory or
correct method name
E/S
MEASURE - ERROR 2:
Faulty method
Programming error on
”OP.SEQUENCE” or
”SEGMENTS” page
−
Check programs on the two pages
and correct
−
If the programming error is not found,
press the softkey
(method test) on the ”MONITORING”
page and note error messages in
error window
−
Correct programs in accordance with
the error messages
E/S
MEASURE - ERROR 3:
Missing sample
−
With automatic program call
Auto.samples = on
, no
sample available in sample
memory
−
With manual sample call
Auto.samples = off
, no
value for
S0
available
−
Enter new sample data or set
Auto.samples
to
off
−
Enter
Sample size/S0
on the
”MONITORING” page
E/S
MEASURE - ERROR 4:
Too many sweeps
The maximum of 60 sweep
blocks has been exceeded
Reduce number of sweeps
E/S