5.3.2 STAND

746 VA Trace Analyzer / 747 VA Stand

5-23

Softkeys

Meaning

695 Auto

sampler

Switch to 695 Autosampler and VA Stand 3

For operation of a 695 Autosampler with the 746 VA Trace Analyzer, a
747 VA Stand with stand address

3

is needed. This softkey is used to

switch to the 695 Autosampler and to the VA Stand with address

3

.

The stand currently selected is displayed in the title line:

Manual Address 1

VA Stand with address 1

Manual Address 2

VA Stand with address 2

Automatic Address 3

VA Stand for Autosampler with address 3

717 Smpl

Changer

Switch to 717 Sample Changer

Reserved for future applications. This softkey will be used to switch to
the initialization and manual operation of the 717 Sample Changer via
746 VA Trace Analyzer.

Rinsing

Dosinos

Switch to rinsing Dosinos

The measuring vessel at the 747 VA Stand can be rinsed by means of
two 700 Dosinos with 50 mL dosing units screwed to supply contain-
ers, one of which adds the rinsing solution while the other siphons off
the wash solution into the waste container. The rinsing Dosinos must
be attached to the 746 VA Trace Analyzer and initialized according to

section 3.10.5. This softkey is used to switch to the manual operation
of the two rinsing Dosinos. The following functions are available:

Rinse

meas.v.

Rinse measuring vessel

After this softkey has been pressed, the measuring vessel is auto-
matically rinsed as follows:

The solution in the measuring vessel is siphoned off.

The measuring vessel is filled with the volume of rinsing solu-
tion defined as

V.rinse

.

The rinsing solution is siphoned off.

The measuring vessel is filled with the volume of rinsing solu-
tion defined as

V.standby

(conditioning solution).

Prime

Dosinos

Initialize rinsing Dosinos

The burette cylinders of the rinsing Dosinos and all connecting tubing
up to the measuring vessel are automatically filled with rinsing solu-
tion.

Empty

Dosinos

Empty rinsing Dosinos

The burette cylinders of the rinsing Dosinos and all connecting tubing
up to the measuring vessel are emptied.

Exchange

position

Switch rinsing Dosinos to exchange position

The current position of the two rinsing Dosinos is shown in the pa-
rameter

Rinse unit pos.

:

exch.

Exchange position (dosing unit can be changed)

rinse

Rinse position

To fill or empty the containers or to remove the dosing units from the
Dosinos, the Dosinos must be in the

exch.

position. If the Dosinos

are in the

rinse

position, they have to be switched to the exchange

position by pressing

.

2