3.13 Connection of devices via remote interface

746 VA Trace Analyzer / 747 VA Stand

3-57

0
I

Control Lines

Printer

External Bus

Dos. 1

Dos. 2

Dos. 3

Dos. 4

RS 232 Interface 1

RS 232 Interface 2

Reset

2A(TH)

115V

230V

230V

746

Fig. 25: Connection of external devices to the remote interface

746 VA Trace Analyzer

external

Inputs

+5V

Pin

Func-

tion
21

input 1

9

input 2

22

input 3

10

input 4

23

input 5

11

input 6

24

input 7

12

input 8

Voltage

Scan

at input

status

0 V

1 (on)

5 V

0 (off)

Outputs

Pin

Func-

tion
5

output 1

18

output 2

4

output 3

17

output 4

3

output 5

16

output 6

1

output 7

2

output 8

External device

V

CEO

40 V

IC

20 mA

Voltage

Scan

at input

status

0 V

1 (on)

open

0 (off)

Voltages

+5V

0V

Pin

Function

15

5 V: inactive, High (H)

14

0 V: active, Low (L)

25

0 V: active, Low (L)

I

40 mA

R

i

12

Contact arrangement at socket
"Control Lines" (female)

13

1

25

14

Connecting cables to external devices must have an
appropriate 25-pin connector (male). Metrohm offers
connector sockets (K.210.9060) and the associated
housing (K.210.0002) as an option for the prepara-
tion of such cables.