13 connection of devices via remote interface – Metrohm 746 VA Trace Analyzer User Manual
Page 84

3.13 Connection of devices via remote interface
746 VA Trace Analyzer / 747 VA Stand
3-57
0
I
Control Lines
Printer
External Bus
Dos. 1
Dos. 2
Dos. 3
Dos. 4
RS 232 Interface 1
RS 232 Interface 2
Reset
2A(TH)
115V
230V
230V
746
Fig. 25: Connection of external devices to the remote interface
746 VA Trace Analyzer
external
Inputs
+5V
Pin
Func-
tion
21
input 1
9
input 2
22
input 3
10
input 4
23
input 5
11
input 6
24
input 7
12
input 8
Voltage
Scan
at input
status
0 V
1 (on)
5 V
0 (off)
Outputs
Pin
Func-
tion
5
output 1
18
output 2
4
output 3
17
output 4
3
output 5
16
output 6
1
output 7
2
output 8
External device
V
CEO
≤
40 V
IC
≤
20 mA
Voltage
Scan
at input
status
0 V
1 (on)
open
0 (off)
Voltages
+5V
0V
Pin
Function
15
5 V: inactive, High (H)
14
0 V: active, Low (L)
25
0 V: active, Low (L)
I
≤
40 mA
R
i
≅
12
Ω
Contact arrangement at socket
"Control Lines" (female)
13
1
25
14
Connecting cables to external devices must have an
appropriate 25-pin connector (male). Metrohm offers
connector sockets (K.210.9060) and the associated
housing (K.210.0002) as an option for the prepara-
tion of such cables.
No liability whatsoever will be accepted for damage arising from improper connection of devices.
Remote cable
External device