Metrohm 746 VA Trace Analyzer User Manual
Page 130

4.4 Lead determination with internal standard
746 VA Trace Analyzer / 747 VA Stand
4-45
2
Switch to dialog page "SEGMENT"
•
Press the yellow
appears with the program for measured value acquisition. The segment
DPASV
is virtually identical to the segment of the same name of the
CALCRV.mth
method (explanation, see section 4.3.3), except the enrich-
ment and start voltage is –800 mV instead of –600 mV.
3
Switch to dialog page "SUBSTANCES"
•
Press the yellow
CES" appears with the parameters for evaluation of the substance lead.
Except for the characteristic voltage
U.verify
(–400 mV), these are
identical to those of the
TESTPb.mth
method and are not explained fur-
ther here (see section 4.2.3).
•
Press the softkey
< >> >
. The subpage of "SUBSTANCES" appears with
the parameters for calibration of the substance lead:
Method: INTSTD SUBSTANCES
Pb - DPASV
Calibration 1996-11-22 16:35 Coefficients
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДДДДДДДД
Technique int std. Prop.fact. 0.5699
Mode analyte Mean dev. 0.03553
Ref.subst. CdRef
Temperature 20.0 °C
# CSoln.name Mass conc.
Range min g/L ДДДДДДДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДДДДДДДД
Range max g/L 0 Pb-1 2.5 ug/L
1 Pb-2 5 ug/L
M.conc./cm g/L 2 Pb-3 10 ug/L
M.conc.begin zero 3 Pb-4 20 ug/L
4 g/L
V0.CSoln. 10.000 mL 5 g/L
6 g/L
7 g/L
8 g/L
9 g/L
Modify
column
Modify
line
Select
subst.
Link to
segment
Delete
subst.
Subst.
name
>>
•
The following is a brief explanation of the most important parameters:
Technique
Calibration technique: internal standard
Mode
Mode for parameter display: analyte
Ref.subst.
Name of reference substance:
CdRef
Temperature
Temperature of the analysis solution: 20°C
V0.CSoln.
Volume added of every calibration solution: 10 mL
Prop.fact.
Proportionality factor between analyte and reference
substance
Mean dev.
Mean deviation of the calibration curve
CSoln.name
Name of the calibration solutions
Mass conc.
Mass concentration of the calibration solutions