5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-204

CAL)n

Page:

O, S

Stand:

M, A

Abbreviation:

C

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. CAL)n .....

Function

End of the loop for recording the calibration curve.

Parameter

Range; Default value / Meaning

n

0...9; 0

Number of calibration solutions – 1

This parameter defines the number of times the CAL loop is executed,
i.e. how many calibration solutions are used to determine the calibra-
tion curve.

Description

The

CAL)n

instruction ends a loop opened with

(CAL

for the recording

of a calibration curve with the calibration technique with calibration curve

cal.crv.

or with internal standard

int.std.

.

In calibrations with a calibration curve, for the recording of this curve

the CAL loop must include one of the dosing instructions

CSOLN/M

,

CSOLN>M

or

CSOLN/F

for addition of the calibration solutions as well as a

sweep or segment call-up. In addition, any other instructions can be
incorporated.

In calibrations with an internal standard, for the recording of the

calibration curve to determine the proportionality factor the CAL loop
must include one of the dosing instructions

CSOLN/M

,

CSOLN>M

or

CSOLN/F

for addition of the calibration solutions, the instruction

RADD>M

or

RADD>F

for addition of the internal standard as well as a sweep or

segment call-up. In addition, any other instructions can be incorporated.

Comments

see

(CAL

Examples

see

(CAL

see also

(CAL, CSOLN/M, CSOLN>M, CSOLN/F, EXIT,
Calibration curve (section 5.6.3 and section 6.9.5),
Internal standard (section 5.6.3 and section 6.9.6)