204 cal)n, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 355

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-204
CAL)n
Page:
O, S
Stand:
M, A
Abbreviation:
C
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. CAL)n .....
Function
End of the loop for recording the calibration curve.
Parameter
Range; Default value / Meaning
n
0...9; 0
Number of calibration solutions – 1
This parameter defines the number of times the CAL loop is executed,
i.e. how many calibration solutions are used to determine the calibra-
tion curve.
Description
The
CAL)n
instruction ends a loop opened with
(CAL
for the recording
of a calibration curve with the calibration technique with calibration curve
cal.crv.
or with internal standard
int.std.
.
In calibrations with a calibration curve, for the recording of this curve
the CAL loop must include one of the dosing instructions
CSOLN/M
,
CSOLN>M
or
CSOLN/F
for addition of the calibration solutions as well as a
sweep or segment call-up. In addition, any other instructions can be
incorporated.
In calibrations with an internal standard, for the recording of the
calibration curve to determine the proportionality factor the CAL loop
must include one of the dosing instructions
CSOLN/M
,
CSOLN>M
or
CSOLN/F
for addition of the calibration solutions, the instruction
RADD>M
or
RADD>F
for addition of the internal standard as well as a sweep or
segment call-up. In addition, any other instructions can be incorporated.
Comments
see
(CAL
Examples
see
(CAL
see also
(CAL, CSOLN/M, CSOLN>M, CSOLN/F, EXIT,
Calibration curve (section 5.6.3 and section 6.9.5),
Internal standard (section 5.6.3 and section 6.9.6)