7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-54

7.6.7

Checking the VA Stand

(performing measurement routine with program simulator)

1 Prepare instrument for diagnostic test (see section 7.6.2).

2

3

press several times and check the on/off switching of the valve

4

press several times and check the on/off switching of the valve

5

press several times and check the on/switching of the motor

The VA Stand includes a dummy cell and a polarogram simulator. In polarography,

this polarogram simulator replaces the electrochemical system (sample, electrode)

and thus performs extremely valuable service if a malfunction occurs to allow the

user to decide whether the malfunction is traceable to the instrument or the electro-

chemical system. This simulator can naturally also be used to practice operating

the instrument.
In the factory a special test method is stored in the VA Stand. This test method

TEST747.mth

is written for the polarogram simulator and allows virtually 100% con-

trol of the entire measurement system.
Should the test method be deleted by accident, a new one can be generated with

the aid of the enclosed program sequences.

6 Copy test method

TEST747.mth

to working storage.

Position cursor on

TEST747.mth

.

,

7 Define sample size.

Position cursor on

Sample size/S0

.

<

*

>,

8 Start

TEST747.mth

method.

The message

AE->AE / RE->RE

appears on the screen.

Attach electrode cable AE to clamping screw AE.

Attach electrode cable RE to clamping screw RE.

The message

WE -> WE-L

appears on the screen.

Attach electrode cable WE to clamping screw WE-L.

The message

WE -> WE-D

appears on the screen.