7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-54
7.6.7
Checking the VA Stand
(performing measurement routine with program simulator)
1 Prepare instrument for diagnostic test (see section 7.6.2).
2
3
press several times and check the on/off switching of the valve
4
press several times and check the on/off switching of the valve
5
press several times and check the on/switching of the motor
The VA Stand includes a dummy cell and a polarogram simulator. In polarography,
this polarogram simulator replaces the electrochemical system (sample, electrode)
and thus performs extremely valuable service if a malfunction occurs to allow the
user to decide whether the malfunction is traceable to the instrument or the electro-
chemical system. This simulator can naturally also be used to practice operating
the instrument.
In the factory a special test method is stored in the VA Stand. This test method
TEST747.mth
is written for the polarogram simulator and allows virtually 100% con-
trol of the entire measurement system.
Should the test method be deleted by accident, a new one can be generated with
the aid of the enclosed program sequences.
6 Copy test method
TEST747.mth
to working storage.
•
•
Position cursor on
TEST747.mth
.
•
,
7 Define sample size.
•
•
Position cursor on
Sample size/S0
.
•
•
<
*
>,
8 Start
TEST747.mth
method.
•
The message
AE->AE / RE->RE
appears on the screen.
•
Attach electrode cable AE to clamping screw AE.
•
Attach electrode cable RE to clamping screw RE.
•
The message
WE -> WE-L
appears on the screen.
•
Attach electrode cable WE to clamping screw WE-L.
•
The message
WE -> WE-D
appears on the screen.