238 f>m, O, s – Metrohm 746 VA Trace Analyzer User Manual

Page 389

background image

5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-238

F>M

Page:

O, S

Stand:

A

Abbreviation:

F>M

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. F>M ..... addition FEED -> MEAS V.add ........mL

Function

Transfer a solution from the sample vessel (FEED) on the 695 Autosam-
pler to the measuring vessel (MEAS) on the 747 VA Stand.

Parameter

Range; Default value / Meaning

V.add

0.010...50.000 mL; 1.000 mL

Transfer volume

Solution volume that should be transferred from the sample vessel on
the sample rack of the 695 Autosampler to the measuring vessel on the
747 VA Stand.

Description

The

F>M

instruction transfers the solution volume specified under

V.add

from the sample vessel (FEED) on the 695 Autosampler to the measur-
ing vessel (MEAS) on the 747 VA Stand.

Comments

The

F>M

instruction can be used only in operation with the 695

Autosampler.

If the solution volume in the sample vessel <

V.add

, an error

message appears.

see also

A>M

,

F>A