4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-40

8

Printout of reports and curves

On completion of the program, the reports and curves entered on the

"DOCUMENTATION" page are put out on the printer:

================== METROHM 746 VA TRACE ANALYZER (5.746.0101) ==================

Determ. : 12160827 User: Date: 1996-12-16

Modified : no Run : 3 Time: 08:27:07

Sample table: -

--------------------------------------------------------------------------------

Pos. Ident.1/S1 Ident.2/S2 Ident.3/S3 Method.call Sample size/S0

Drink water 10 mL

--------------------------------------------------------------------------------

Method : CALCRV

Title : Pb determination with calibration curve

Remark1 :

Remark2 :

--------------------------------------------------------------------------------

Substance : Pb Comments

Mass conc.: 12.48 ug/L Mass : 124.8 ng ––––––––––––––––––

MC.dev. : 0.344 ug/L (2.76%) Add.mass : - dev. estimated

Cal.dev. : 0.195 ug/L (1.57%) V0.sample: 10 mL

VR U/mV I/nA I.mean Std.dev. I.delta Comments

––– ––––– –––––– –––––– ––––––– ––––––– ––––––––––––––––––

00 -409 16.57 16.86 0.2620

01 -409 16.96

02 -410 17.06

Substance Techn. Y.reg/offset Slope Nonlin. Mean deviat.

––––––––– –––––––– –––––––––––– –––––––––– –––––––––– ––––––––––––

Pb cal.crv. 5.489e-10 0.001438 5.933e-10

C# Workg.com.var Remark

–– ––––––––––––– ––––––––––––––––

Final results +/- Res.dev. % Comments

––––––––––––––––––––––––––––––––– ––––––––––––––––– ––––––––––––––––––

Pb = 12.48 ug/L 0.344 2.76 see Evaluations

The full report

Full report

essentially contains the same parts as that

of the

TESTPb.mth

method with standard addition (see section 4.2.6).

However, the following sections are different:

1

Data for substance

Substance name, mass concentration

Mass

evaluation

conc.

found and its absolute and relative total

deviation

MC.dev., the deviation due to the

calibration

Cal.dev., substance mass Mass

found in the measuring vessel and sample volume
V0.sample in the measuring vessel.

2

Data for calibration

Offset

Y.offset and slope Slope of the

curve

calibration curve

used for the calculation.

3

Final results

Final result with formula calculation.

3

1

2