277 sdos>a, O, s – Metrohm 746 VA Trace Analyzer User Manual
Page 428

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-277
SDOS>A
Page:
O, S
Stand:
A
Abbreviation:
SD
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SDOS>A ..... Soln.name ........ Prop.factor ........
Function
Dispense auxiliary solution proportional to the sample size
S0
into the
auxiliary vessel (AUX) on the 695 Autosampler.
Parameter
Range; Default value / Meaning
Soln.name
8 characters; -
Name of the auxiliary solution
The entry of a solution name is mandatory when the auxiliary solution
should be added automatically. In this case the solution name
Soln.name
must be entered in the parameter of the same name on the
"SOLUTIONS" page and assigned a position
Pos.
(see section 5.3.4).
Without entry of a name for
Soln.name
in the
SDOS>A
instruction,
without entry of this name on the "SOLUTIONS" page or without a
position assignment, the auxiliary solution must be dispensed manually
(the prompt
DOS "Soln.name" "V.add" > A
then appears during the
program run).
Prop.factor
0.0010...99999.9; 1
Proportionality factor
Specification of the proportionality factor for calculation of the addition
volume
V.add
:
V.add (mL) = Prop.factor * S0 (mL or g)
Description
The
SDOS>A
instruction is used to dispense an auxiliary solution (e.g.
supporting electrolyte, reagent solution) proportional to the sample size
S0
into the auxiliary vessel (AUX) on the 695 Autosampler.
For automatic addition with the 695 Autosampler, the appropriate
position of the auxiliary solution must be specified on the "SOLUTIONS"
page (see
Soln.name
).
Comments
The
SDOS>A
instruction can be used only in operation with the 695
Autosampler.
see also
SDOS>M, SDOS>F, DOS/M, DOS>M, DOS>F, DOS>A, PDOS>M