7 safety, errors, troubleshooting, diagnosis, glp – Metrohm 746 VA Trace Analyzer User Manual
Page 519

7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-20
Error message
Meaning/Causes
Rectification
Exit
OP.SEQ. - ERROR 3:
Incorrect bracket
matching: line ##
Wrong use of instructions with
brackets (e.g.
(ADD .....
REP)
)
Check program on the
”OP.SEQUENCE” page for correct use
of instructions with brackets and correct
if necessary
Esc
OP.SEQ. - ERROR 4:
Multiple use of ....
−
.... = (PREP
The
(PREP
or
PREP)
in-
struction is admissible once
only
−
.... = SWEEP
Only one
SWEEP
instruction is
admissible in a segment
−
Delete unnecessary
(PREP
or
PREP)
instructions
−
Delete unnecessary
SWEEP
instruc-
tions
Esc
OP.SEQ. - ERROR 5:
Instruction not allowed
in block: line ##
There is an inadmissible
instruction in the specified
program line
Delete inadmissible instruction on the
”OP.SEQUENCE” page
Esc
OP.SEQ. - ERROR 6:
Missing part of prep.
(PREP, PREP), >PREP
One of the three instructions for
a PREP block missing
In program on the ”OP.SEQUENCE”
page, check that all three instructions
(PREP
,
PREP)
and
>PREP
are present
Esc
OP.SEQ. - ERROR 7:
Not allowed in PREP
block: line ##
Inadmissible instruction in the
PREP block
Delete inadmissible instruction on the
”OP.SEQUENCE” page
Esc
OP.SEQ. - ERROR 8:
Invalid with VARNI:
line ##
Inadmissible parameter for
parameter variation
Check parameter name for
Param.name
on the ”OP.SEQUENCE”
page and correct
Esc
OP.SEQ. - ERROR 9:
Invalid with (CSWEEP:
line ##
Inadmissible instruction in the
CSWEEP block (only
RAMP
is
allowed)
Delete inadmissible instruction in the
CSWEEP block on the
”OP.SEQUENCE” page
Esc
OP.SEQ. - ERROR 10:
Inconsist. value in
.....: line ##
Inadmissible value for the
specified parameter
Change parameter value on the
”OP.SEQUENCE” page
Esc
OP.SEQ. - ERROR 11:
Value inconsist. with
electrode: t.step in
line ##
t.step
value for specified
electrode too low
−
Increase
t.step
or reduce
t.modul
−
Change electrode type
Esc
OP.SEQ. - ERROR 12:
Missing value in line
##
Missing parameter value in the
specified program line
Enter parameter value on the
”OP.SEQUENCE” page
Esc
OP.SEQ. - ERROR 13:
Not allowed data-
generating: line ##
Inadmissible data-generating
instruction in the specified
program line
Delete inadmissible instruction on the
”OP.SEQUENCE” page
Esc
OP.SEQ. - ERROR 14:
Not allowed in segment:
line ##
Inadmissible instruction in the
segment
Delete inadmissible instruction
Esc
OP.SEQ. - ERROR 15:
Segment ........ not
found
The segment with the specified
name is not available
Check name of segment
Segm.name
to
be called in the ”
SEGMENT
” instruction
and correct
Esc
OP.SEQ. - ERROR 16:
Incorrect use of (CAL
... RINSE sequence
If several CAL loops are used in
a program, every loop must
have a
RINSE
instruction
Enter
RINSE
instructions on the
”OP.SEQUENCE” page
Esc
OP.SEQ. - ERROR 17:
Sweep range too small
Sweep range too small
(< 8 measured values)
Increase sweep range or decrease
U.step
or
t.step
Esc
OP.SEQ. - ERROR 18:
Sweep range too large
Sweep range too large
(> 1000 measured values)
Decrease sweep range or increase
U.step
or
t.step
Esc
OP.SEQ. - ERROR 19:
Missing "t.step"
t.step
value missing or is
<1 ms (possible only in
variation loops)
Enter variation loop for
t.step
in the
segment
so that
t.step
remains >1 ms
Esc
OP.SEQ. - ERROR 20:
Conditioning cycle
range too small
Initial voltage
U.start
and
final voltage
U.end
of the
COND
instruction are identical
Change
U.start
and/or
U.end
in the
COND
instruction
Esc