7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-20

Error message

Meaning/Causes

Rectification

Exit

OP.SEQ. - ERROR 3:
Incorrect bracket
matching: line ##

Wrong use of instructions with

brackets (e.g.

(ADD .....

REP)

)

Check program on the

”OP.SEQUENCE” page for correct use

of instructions with brackets and correct

if necessary

Esc

OP.SEQ. - ERROR 4:
Multiple use of ....

.... = (PREP

The

(PREP

or

PREP)

in-

struction is admissible once

only

.... = SWEEP

Only one

SWEEP

instruction is

admissible in a segment

Delete unnecessary

(PREP

or

PREP)

instructions

Delete unnecessary

SWEEP

instruc-

tions

Esc

OP.SEQ. - ERROR 5:
Instruction not allowed
in block: line ##

There is an inadmissible

instruction in the specified

program line

Delete inadmissible instruction on the

”OP.SEQUENCE” page

Esc

OP.SEQ. - ERROR 6:
Missing part of prep.
(PREP, PREP), >PREP

One of the three instructions for

a PREP block missing

In program on the ”OP.SEQUENCE”

page, check that all three instructions

(PREP

,

PREP)

and

>PREP

are present

Esc

OP.SEQ. - ERROR 7:
Not allowed in PREP
block: line ##

Inadmissible instruction in the

PREP block

Delete inadmissible instruction on the

”OP.SEQUENCE” page

Esc

OP.SEQ. - ERROR 8:
Invalid with VARNI:
line ##

Inadmissible parameter for

parameter variation

Check parameter name for

Param.name

on the ”OP.SEQUENCE”

page and correct

Esc

OP.SEQ. - ERROR 9:
Invalid with (CSWEEP:
line ##

Inadmissible instruction in the

CSWEEP block (only

RAMP

is

allowed)

Delete inadmissible instruction in the

CSWEEP block on the

”OP.SEQUENCE” page

Esc

OP.SEQ. - ERROR 10:
Inconsist. value in
.....: line ##

Inadmissible value for the

specified parameter

Change parameter value on the

”OP.SEQUENCE” page

Esc

OP.SEQ. - ERROR 11:
Value inconsist. with
electrode: t.step in
line ##

t.step

value for specified

electrode too low

Increase

t.step

or reduce

t.modul

Change electrode type

Esc

OP.SEQ. - ERROR 12:
Missing value in line
##

Missing parameter value in the

specified program line

Enter parameter value on the

”OP.SEQUENCE” page

Esc

OP.SEQ. - ERROR 13:
Not allowed data-
generating: line ##

Inadmissible data-generating

instruction in the specified

program line

Delete inadmissible instruction on the

”OP.SEQUENCE” page

Esc

OP.SEQ. - ERROR 14:
Not allowed in segment:
line ##

Inadmissible instruction in the

segment

Delete inadmissible instruction

Esc

OP.SEQ. - ERROR 15:
Segment ........ not
found

The segment with the specified

name is not available

Check name of segment

Segm.name

to

be called in the ”

SEGMENT

” instruction

and correct

Esc

OP.SEQ. - ERROR 16:
Incorrect use of (CAL
... RINSE sequence

If several CAL loops are used in

a program, every loop must

have a

RINSE

instruction

Enter

RINSE

instructions on the

”OP.SEQUENCE” page

Esc

OP.SEQ. - ERROR 17:
Sweep range too small

Sweep range too small

(< 8 measured values)

Increase sweep range or decrease

U.step

or

t.step

Esc

OP.SEQ. - ERROR 18:
Sweep range too large

Sweep range too large

(> 1000 measured values)

Decrease sweep range or increase

U.step

or

t.step

Esc

OP.SEQ. - ERROR 19:
Missing "t.step"

t.step

value missing or is

<1 ms (possible only in

variation loops)

Enter variation loop for

t.step

in the

segment

so that

t.step

remains >1 ms

Esc

OP.SEQ. - ERROR 20:
Conditioning cycle
range too small

Initial voltage

U.start

and

final voltage

U.end

of the

COND

instruction are identical

Change

U.start

and/or

U.end

in the

COND

instruction

Esc