Metrohm 746 VA Trace Analyzer User Manual
Page 267

5.5 Operation
746 VA Trace Analyzer / 747 VA Stand
5-116
Cyclic Segment-VR-Curve: XXXXXXXX-XX-original
Select
segment
Start
zoom
Reset
zoom
Next
VR
Prev.
VR
Scale
Softkeys
Meaning
Select
segment
Select segment curves
The selection of segment and VR numbers determines what cyclic
curves are displayed. A cyclic segment curve always includes all of
the maximum 5 ramps which have the same VR number (the display
of individual ramps is not possible). First, the softkeys carrying the
segment names must be used to select the desired cyclic segment:
Segment#
Selection of the segment
Segment#
.
This is followed by entry of the numbers for variation
V
and replication
R
. For this, the system window shows the message:
VR max: ##
Maximum selectable VR values
Enter VR: [0*]
Entry of VR
The following possibilities are available for the entry of VR:
##
Entry of two numbers
##
for the display of a single cy-
clic curve with all ramps of the same measurement cy-
cle (measurement cycles = replications).
*#
Entry of an asterisk
*
and a number
#
for the display
of all variations with the same replication (shown by
(grouped VR: *#)
in the title line).
#*
Entry of a number
#
and an asterisk
*
for the display
of all cyclic curves of the same variation (shown by
(grouped VR: #*)
in the title line.
**
Entry of two asterisks
**
for the display of all cyclic
curves (shown by
(grouped VR: **)
in the title line).
Without entry of VR, the default value
0*
is effective for variation 0
with all replications as standard.
15