beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 253

background image

5.5 Operation

746 VA Trace Analyzer / 747 VA Stand

5-102

Parameter

Range; Default value / Meaning

dU.rear

display only

Voltage difference to rear base point

Absolute voltage difference determined between the base point at the
peak/wave rear half and the peak voltage

U.peak

or the half-wave

potential

U/2

in mV.

S.front

display only

Slope of baseline at front base point

Slope of the baseline determined at the base point at the peak/wave
front half.

With peaks with a baseline of the

linear

type, this parameter is not

shown.

S.rear

display only

Slope of baseline at rear base point

Slope of the baseline determined at the base point at the peak/wave
front half.

With peaks with a baseline of the

linear

type, this parameter is not

shown.

linear, whole

dU.frontdU.rear

S.rear

S.front

S.rear

S.front

S.front

S.rear

linear, DCT

dU.frontdU.rear

expon., whole

dU.frontdU.rear

polynom, whole

dU.frontdU.rear