5.6.3 SUBSTANCES

746 VA Trace Analyzer / 747 VA Stand

5-157

Parameter

Range; Default value / Meaning

V0.CSoln.

0.1...50.0 mL; 5.000 mL

Addition volume of calibration solutions

Volume of every calibration solution that must be added to the meas-
uring vessel.
If work is performed with the 695 Autosampler, an asterisk

*

can be

entered here in place of a volume. However, this is admissible only if
the instruction

CSOLN/F

with subsequent solution transfer to the

auxiliary or measuring vessel is used.
If calibration curves of several substances are recorded simultane-
ously,

V0.CSoln.

must be the same for every substance.

Coefficients

display only

Coefficients

Title for the calculated coefficients of the calibration curve for the
internal standard.

Prop.fact.

-999.9...999.9 e-20...e+20;

Proportionality factor

Display of the proportionality factor between analyte and reference
substance. The proportionality factor is identical with the slope calcu-
lated from the calibration lines. This value can also be entered or
changed manually.
This value can be used in formulae as

KS:

(

= sub-

stance name) for further calculations.

Mean dev.

-999.9...999.9 e-20...e+20;

Mean deviation of calibration curve for internal standard

Display of the calculated, mean deviation of the measured values
about the calibration curve. This value can be changed manually.
The size of

Mean dev.

can be used as a measure of the quality of the

calibration curve for the internal standard.
This value can be used in formulae as

KD:

(

= sub-

stance name) for further calculations.

CSoln.name

8 characters;

Name of calibration solutions

The entry of a name for the calibration solutions is optional. The
names are used as follows:

Without 695 Autosampler, before every manual addition of a
calibration solution to the measuring vessel the message

Next

Standard, CSOLN 'CSOLN.name''Mass conc.' > M

appears with

details of the name and the mass concentration of the calibration
solution.

If work is performed with the 695 Autosampler with

Auto.samples

= on

, all calibration solutions must be entered on the "SAMPLES"

page. We advise you to enter the names

CSoln.name

there as

sample identification.

Mass conc.

0.001...999.99 ng...kg/L;

Mass concentration

ρρ

a,n

of calibration solutions

The mass concentration must be specified for all n calibration solu-
tions defined by the instruction

CAL)n

.