5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-311

VWASH

Page:

O, S

Stand:

A

Abbreviation:

VW

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. VWASH ..... V.add ........mL

Function

Additional wash volume for subsequent dispensing or transfer instruction
on the 695 Autosampler.

Parameter

Range; Default value / Meaning

V.add

0.010...10.000 mL; 1.000 mL

Addition volume

Additional solution volume that is delivered following the dispensing or
transfer instruction.

Description

The

VWASH

instruction delivers an additional wash volume

V.add

of

hydraulic solution following a dispensing or transfer instruction on the
695 Autosampler. This ensures that the preceding solution transfer was
actually complete.

Comment

The

VWASH

instruction refers only to the immediately following

dispensing or transfer instruction.

The

VWASH

instruction can be used with the following dispensing or

transfer instructions for the 695 Autosampler:

SMPL>M
DOS>M, DOS>A, DOS>F, SDOS>M, SDOS>A, SDOS>F, PDOS>M
ADD>M, PADD>M, CSOLN>M, RADD>M, RADD>F
A>M, F>A, F>M

For solution transfers on the 695 Autosampler, the following generally
holds:

Transfer volume

VWASH

Siphoning off Delivery

< 200

µ

L

V.add

= 0

100%

100%

V.add

> 0

100%

100% +

V.add

200

µ

L

V.add

= 0

105%

100% (+ 5% in waste)

V.add

> 0

100%

100% +

V.add

see also

DOS>A, DOS/M, DOS>M, DOS>F, PDOS>M, SDOS>M, SDOS>A,

SDOS>F, ADD/M, ADD>M, PADD>M, RADD>F, RADD>M