311 vwash, O, s – Metrohm 746 VA Trace Analyzer User Manual
Page 462

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-311
VWASH
Page:
O, S
Stand:
A
Abbreviation:
VW
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. VWASH ..... V.add ........mL
Function
Additional wash volume for subsequent dispensing or transfer instruction
on the 695 Autosampler.
Parameter
Range; Default value / Meaning
V.add
0.010...10.000 mL; 1.000 mL
Addition volume
Additional solution volume that is delivered following the dispensing or
transfer instruction.
Description
The
VWASH
instruction delivers an additional wash volume
V.add
of
hydraulic solution following a dispensing or transfer instruction on the
695 Autosampler. This ensures that the preceding solution transfer was
actually complete.
Comment
•
The
VWASH
instruction refers only to the immediately following
dispensing or transfer instruction.
•
The
VWASH
instruction can be used with the following dispensing or
transfer instructions for the 695 Autosampler:
SMPL>M
DOS>M, DOS>A, DOS>F, SDOS>M, SDOS>A, SDOS>F, PDOS>M
ADD>M, PADD>M, CSOLN>M, RADD>M, RADD>F
A>M, F>A, F>M
•
For solution transfers on the 695 Autosampler, the following generally
holds:
Transfer volume
VWASH
Siphoning off Delivery
< 200
µ
L
V.add
= 0
100%
100%
V.add
> 0
100%
100% +
V.add
≥
200
µ
L
V.add
= 0
105%
100% (+ 5% in waste)
V.add
> 0
100%
100% +
V.add
see also
DOS>A, DOS/M, DOS>M, DOS>F, PDOS>M, SDOS>M, SDOS>A,