5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-248

ØMEAS

Page:

O, S

Stand:

M, A

Abbreviation:

Ø

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. ØMEAS ..... U.standby ........mV

Function

Switch off voltage, current and DME.

Parameter

Range; Default value / Meaning

U.standby

-3000...3000 mV; -

Standby voltage after measurement

The voltage entered here is applied to the electrodes. This voltage
remains in force even after the end of the main program on the
"OP.SEQUENCE" page. It is switched off only by a new

ØMEAS

instruc-

tion with

U.standby

= empty, by a

MEAS

instruction or by the new start

of a method.

The standby voltage

U.standby

is displayed on the "STAND" page.

If the voltage applied to the electrodes should be switched off,

U.standby

must be empty (press to empty the contents of

the field).

Description

The

ØMEAS

instruction is used to switch off the current measurement and

to apply the standby voltage

U.standby

to the electrodes. The dropping

mercury electrode DME is also switched off, i.e. the mercury flow is
stopped.

Comment

If the solution in the measuring vessel is changed during the program
run (e.g. in the recording of calibration curves), the

ØMEAS

instruction with

no standby voltage set (

U.standby

= empty) must be inserted before the

solution change so that no voltage flows through the electrodes.

Examples

see

(CAL

see also

MEAS