5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-248
ØMEAS
Page:
O, S
Stand:
M, A
Abbreviation:
Ø
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. ØMEAS ..... U.standby ........mV
Function
Switch off voltage, current and DME.
Parameter
Range; Default value / Meaning
U.standby
-3000...3000 mV; -
Standby voltage after measurement
The voltage entered here is applied to the electrodes. This voltage
remains in force even after the end of the main program on the
"OP.SEQUENCE" page. It is switched off only by a new
ØMEAS
instruc-
tion with
U.standby
= empty, by a
MEAS
instruction or by the new start
of a method.
The standby voltage
U.standby
is displayed on the "STAND" page.
If the voltage applied to the electrodes should be switched off,
U.standby
must be empty (press
the field).
Description
The
ØMEAS
instruction is used to switch off the current measurement and
to apply the standby voltage
U.standby
to the electrodes. The dropping
mercury electrode DME is also switched off, i.e. the mercury flow is
stopped.
Comment
If the solution in the measuring vessel is changed during the program
run (e.g. in the recording of calibration curves), the
ØMEAS
instruction with
no standby voltage set (
U.standby
= empty) must be inserted before the
solution change so that no voltage flows through the electrodes.
Examples
see
(CAL
see also
MEAS