6.9 Content calculation

746 VA Trace Analyzer / 747 VA Stand

6-31

working range on the dialog page "SUBSTANCES" with the two parameters

Range min

and

Range max

. If a mass concentration

Mass conc.

outside the

defined working range is found for the substance, an error message is put

out and

conc. out of range

displayed on the results page.

Keep temperature constant

Owing to the large temperature dependence of the measured values

(

2%/°C), it is advisable to work with the 6.1418.220 thermostatted measur-

ing vessel.

Consideration of blank values

Any blank values must be determined separately and subtracted in the for-

mula calculation on the dialog page "CALCULATIONS".

6.9.6

Internal standard

In the method of internal standards a known quantity of a reference substance (r =

reference) is added to the analysis solution. The mass concentration

ρ

eff,a

eff,a

of the

analyte (a = analyte) is calculated as follows from the ratio of the two evaluation

quantities

EV

a

/

EV

r

and the proportionality factor

Prop.fact.

determined before-

hand by measurement of different reference solutions:

ρ

aa

=

ρ

rr

×

EV

a

/

EV

r

×

1/

Prop.fact.

As in the case of the calibration curve, the content is determined with an internal

standard in two steps:

– First the relation between the ratio of the mass concentrations

ρ

eff,a

eff,a

/

ρ

eff,r

eff,r

of

analyte and reference substance and the ratio of the evaluation quantities

EV

a

/

EV

r

is determined by measurement of different reference solutions.

– Finally, the ratio of the mass concentrations

ρ

eff,a

eff,a

/

ρ

eff,r

eff,r

in the sample is meas-

ured and the mass concentration

ρ

eff,a

eff,a

determined using the recorded cali-

bration curve.

The method of internal standards is programmed on the dialog page "OP.SE-

QUENCE" with the following instructions and parameters (see section 5.7.3):

(CAL...CAL)n

Loop for recording the calibration curve

n

Number of calibration solutions – 1

CSOLN...

Addition of the calibration solutions

RADD...

Addition of the internal standard

Soln.name

Name of the solution with the internal standard

V.add

Addition volume of the solution with the internal standard

The remaining parameters needed for the internal standards method are entered

on the dialog page "SUBSTANCES/Calibration" (see section 5.6.3). The most im-

portant of these are:

Technique

Selection of the calibration technique

int.std.

Internal standard

Mode

Mode for parameter display

analyte

Measured substance (analyte)

referen.

Reference substance (internal standard)