5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-297

STIR

Page:

O, S

Stand:

M, A

Abbreviation:

ST

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. STIR ..... Rot.speed ......../min

Function

Switch on stirrer.

Parameter

Range; Default value / Meaning

Rot.speed

0, 200, 400, ..., 3000 /min; 2000 /min

Rotational speed

Setting of the number of revolutions of the stirrer per min.

Description

The

STIR

instruction is used to switch on the stirrer of the 747 VA Stand.

The stirrer comprises the 6.1246.010 Drive shaft and the 6.1204.090
Stirrer tip (see section 3.8).

The stirrer remains active until it is switched off by the

ØSTIR

instruction

or

END

in the main program on the "OP.SEQUENCE" page.

Comment

If the stirrer should be switched off automatically after a definite time, the

TSTIR

instruction can also be used.

see also