3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-26

6 Replace sealing needle 97

97 if need be

If problems with leaks arise owing to a worn, deformed or damaged sealing

needle 97

97 (see sections 7.4.2 and 7.4.3), this must be replaced. Three spare

needles are supplied separately in a protective plastic package. After un-

packing a needle, please avoid any contact whatsoever with the needle tip.

The spare needle 97

97 is installed as follows:

Carefully pull old sealing needle 97

97 out of PTFE membrane of the slotted

screw 96

96.

Carefully insert new sealing needle 97

97 without tilting into the hole in the

PTFE membrane of the slotted screw 96

96.

When the sealing needle 97

97 is changed, it is always necessary to

change the glass capillary 109

109 !

7 Replace sealing ring 106

106 if need be

If the sealing ring 106

106 is contaminated or damaged in any way (see

sections 7.4.2 and 7.4.3), it must be replaced for the subsequent assem-

bly of the MME. Two new sealing rings 106

106 are enclosed in the package

with the 6.1226.030 glass capillaries.

8 Reassemble MME

Screw slotted screw 101

101 using a suitable coin flush into screw thread

100

100.

Do not turn so tightly that the cemented-in steel threaded ring 100

100

becomes loose and hence jeopardizes the tightness and safety of the

MME!

Using a suitable coin, screw slotted screw 96

96 into the screw thread 98

98

until the contact surface of the black O-ring at the Plexiglas wall (thin,

black stripe) is just visible below the metal thread 98

98.

Place multi-mode electrode 38

38 with the opening 104

104 facing upwards in

the electrode holder 114

114 (see Fig. 15-1).

Insert insert ring 105

105 in opening 104

104.

Push sealing ring 106

106 onto locking ring 107

107 and insert both in the open-

ing 104

104.

Screw retaining nut 108

108 by hand into screw thread 104

104 until a slight

resistance is felt.

9 Add mercury

Proceed as described in section 3.4.2.

10 Mount new capillary

Proceed as described in section 3.4.3.

11 Fill capillary

Proceed as described in section 3.4.4 or section 3.4.5.