Error message

Meaning/Causes

Rectification

Exit

CURVES - ERROR 3:
No segment available

No segment data whatsoever

available

Start new measurement

E

CURVES - ERROR 4:
Selected segment not
found

The selected segment (VR) was

not found

Switch to available segment

E

CURVES - ERROR 5:
Can not execute,
missing data

Action can not be executed

owing to missing result data

Start new measurement

E

CURVES - ERROR 6:
Can not display
calibration curve

No calibration curve available

If really desired, start new measurement

with calibration

E

CURVES - ERROR 7:
Can not handle
meas.data with actual
method

The current method has been

so changed that the experi-

mental data curves can no

longer be displayed

Reload determination with original

method or start new measurement

E

CURVES - ERROR 8:
Unqualified peak
attribute for this
operation

The requested operation can

not be performed with the

selected peak (e.g.

baseline>

with peak without

baseline)

Read remarks on selected peak on the

”RESULTS/Evaluation” page

E

CURVES - ERROR 9:
Can not find plot
parameters of segment

The segment curve can not be

displayed as no plot parame-

ters are available (e.g. as the

segment name has been

changed)

Reset segment name (the name may

not be changed after the determination

has been recorded)

E

DATA CARD - ERROR 1:
Hardware error

Hardware fault on data card or

unexpected write/read error

Check data card, reinsert and attempt

write/read operation again

If error reappears, repeat write/read

test with a different data card (if suc-

cessful, reformat old card)

If error reappears, inform Metrohm

service

E/C

DATA CARD - ERROR 2:
Card invalid or not
available

No data card inserted

Data card not inserted

properly

Battery of data card fully

discharged

Insert data card

Take out data card then reinsert and

push right in

Change battery of data card

E/C

DATA CARD - ERROR 3:
Battery word bad

Test pattern in RAM of data

card destroyed, i.e. data card

battery discharged

Change battery of data card

E/C

DATA CARD - ERROR 4:
Low battery charge
detected

Charge of data card battery low

Copy all data on data card to instru-

ment memory

Change battery

Recopy data to data card

E/C

DATA CARD - ERROR 6:
Invalid file descriptor

Data interchange with data

card not possible

Press Reset key

E/C

DATA CARD - ERROR 7:
Can not open more files
simultaneously

Data interchange with data

card not possible

Press Reset key

E/C

DATA CARD - ERROR 8:
File name already
exists

Error in changing file name:

Inputted new name is inadmis-

sible as a file with the same

name exists (overwrite protec-

tion)

Use a different file name or delete old

file

E/C