Metrohm 746 VA Trace Analyzer User Manual

Page 57

background image

3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-30

3.6.2

Startup procedure

The reference electrode 43

43 is supplied in modular form as the reference system

122

122 and the electrolyte vessel 123

123 and has first to be filled and assembled as fol-

lows:

1 Add internal electrolyte

Filling of the reference system is necessary only when the optional

6.0728.010 Reference system supplied dry is used, if the internal electrolyte

solution has to be renewed or if the electrical connection is interrupted by

gas bubbles.

Hold reference system 122

122 so that diaphragm 129

129 faces upwards.

Unscrew diaphragm support 128

128.

Fill electrolyte compartment 127

127 completely with the desired internal

electrolyte. Expel any air bubbles by tapping shaft gently.

Screw diaphragm support 128

128 back on, the electrolyte solution thus

displaced is expelled through the vent opening 125

125.

2 Add bridge electrolyte

Fill internal compartment 131

131 of the electrolyte vessel 123

123 with a suitable

bridge electrolyte whose composition depends on the analyses to be

performed (aqueous or non-aqueous solution, composition of the sup-

porting electrolyte, etc.).

If you use the same solution for the bridge electrolyte and the internal

electrolyte (single-junction operation), the inner diaphragm 129

129 can

be omitted to reduce the electrical resistance: Unscrew diaphragm

support 128

128 with diaphragm 129

129 from the reference system 122

122.

3 Screw reference electrode together

Insert the filled reference system 122

122 in the vessel 123

123 filled with bridge

electrolyte and screw tight. The electrolyte solution thus displaced is ex-

pelled through the vent opening 130

130.

4 Install reference electrode in 747 VA Stand and connect

Insert reference electrode 43

43 in opening 79

79 of the measuring head 41

41

(see Fig. 9).

Attach electrode cable 48

48 (RE) to reference electrode 43

43:

push cable lug under the screw and then tighten screw firmly.

Turn reference electrode 43

43 so that the electrode cable 48

48 points to the

rear and not to the side (in the latter position it may become kinked and

damaged when cover 24

24 is closed).

Wait approx. 20 min before first measurement (it takes some time for a

new reference electrode to become wet).