Metrohm 746 VA Trace Analyzer User Manual

Page 180

background image

5.3.2 STAND

746 VA Trace Analyzer / 747 VA Stand

5-29

Softkeys

Meaning

Exchange

position

Switch rinsing Dosinos to exchange position

The current position of the two rinsing Dosinos is shown in the pa-
rameter

Rinse unit pos.

:

exch.

Exchange position (dosing unit can be changed)

rinse

Rinse position

To fill or empty the containers or to remove the dosing units from the
Dosinos, the Dosinos must be in the

exch.

position. If the Dosinos

are in the

rinse

position, they have to be switched to the exchange

position by pressing

.

After the solutions or the dosing units have been changed, the Dosi-
nos must be initialized by pressing

. All connecting

tubing up to the measuring vessel and the wash station are automati-
cally filled with rinsing solution.

Electr.

test

Electrode test

The electrode (MME or RDE) used in the 747 VA Stand is tested. The
electrode type must be specified for this:

DME

DME (Dropping Mercury Electrode)

HMDE

HMDE (Hanging Drop Mercury Electrode)

SMDE

SMDE (Static Mercury Drop Electrode)

RDE

RDE (Rotating Disk Electrode)

On successful completion of the test, the message

Electrode test

ok

appears in the system window, if any type of malfunction is found

the appropriate error message is shown (see section 7.3.2).

>>

Rack def

Switch to "Rack definition"

Switches to subpage "Rack definition" for setting the rack positions.

>>

Switches

Switch to "Switches"

Switches the softkey bar to the setting of inert gas pressure, purging,
stirrer and electrodes.