5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-202

The

EXIT

instruction for premature exit from the CAL loop can be

included in this loop.

For the recording of a calibration curve, the program run mode

Run

mode = calibration

must be set on the "MONITORING" page. The

calibration solutions are then measured in turn and the coefficients of
the calibration curve determined. To store these permanently, the
method must then be copied from the working storage to the method
memory (this occurs automatically with

Auto.batch = on

).

Examples

Determinations using a calibration curve are normally programmed on

the "OP.SEQUENCE" page with the following basic structure:

:
C (CAL

Loop start for calibration

C CSOLN>M or CSOLN/M

Addition of calibration solution to

:

measuring vessel

C (REP

Loop start for replication

C SEGMENT Seg1

Call-up of the segment

:

(measurement of the calibration solution)

C REP)n

Loop end, n = number of replications

C ØMEAS

Switch off voltage (solution change)

C RINSE

Reset volume bookkeeping

C CAL)n

Loop end, n = number of calibration solutions – 1

:
D SMPL>M or SMPL/M

Sample addition to measuring vessel

D (REP

Loop start for replication

D SEGMENT Seg1

Call-up of the segment

:

(measurement of the sample)

D REP)n

Loop end, n = number of replications

:

In this example, the instructions for the calibration (

C

) and determination

(

D

) have been clearly separated from each other. If the calibration and

determination use the same number of replications, a shorter version of
the same program can be entered as follows:

:
D SMPL>M or SMPL/M

Sample addition to measuring vessel

C (CAL

Loop start for calibration

C CSOLN>M or CSOLN/M

Addition of calibration solution to

:

measuring vessel

(REP

Loop start for replication

SEGMENT Seg1

Call-up of the segment (measurement of

:

the calibration solution or sample)

REP)n

Loop end, n = number of replications

C ØMEAS

Switch off voltage (solution change)

C RINSE

Reset volume bookkeeping

C CAL)n

Loop end, n = number of calibration solutions – 1

: