5.3 Set Up

746 VA Trace Analyzer / 747 VA Stand

5-44

Softkeys

Meaning

Sort by

name

Sort by name

The solutions are sorted alphabetically by name

Soln.name

.

Sort by

position

Sort by position

The solutions are sorted by their position

Pos.

.

Print

all

Print list of all solutions

The "SOLUTIONS" page is printed out with all solution entries.

Upd.from

method

Takeover solutions from method

The auxiliary solutions defined in the loaded method on the dialog
pages "OP.SEQUENCE", "SEGMENTS" or "SUBSTANCES" are auto-
matically transferred to the solution list. Depending on the place of
origin, the following entries are made in the

Remarks

column:

from:

Solution from subpage "Calibrations" of
the substance with the name

Command from: Op.Seq.

Solution from dispensing instruction on
the "OP.SEQUENCE" page

Command from:

Solution from dispensing instruction in
the segment with the name

Parameter

Range; Default value / Meaning

Soln.name

8 characters;

Solution name

The entry of a solution on this page is required only if this has to be
dispensed automatically with a 685 Dosimat, a 700 Dosino or the 695
Autosampler. In this case, the name entered here must be identical
with the solution name

Soln.name

of the dispensing instruction in

question.

We recommend not to put in the solution names via keypad but to

takeover the solution names defined in the current method by pressing

the softkey

Upd.from method

.

If work is carried out with a single mixed solution for standard addi-
tions of several substances, a separate line with the solution name
and the position of the mixed solution must be entered for each
substance.

Rinsing Dosinos

It is possible to rinse the measuring vessel at the 747 VA Stand manually

via dialog page "STAND" or with the RINSE command by means of two

700 Dosinos with 50 mL dosing units attached to the "Dos. 3" and

"Dos. 4" connections of the 746 VA Trace Analyzer (see section 3.10.5).

To make possible this rinsing, a solution with the solution name

Soln.name = RINSE

(upper case letters) and the position

Pos. = Dos.3

has to be entered.