Metrohm 746 VA Trace Analyzer User Manual
Page 138

4.4 Lead determination with internal standard
746 VA Trace Analyzer / 747 VA Stand
4-53
•
The status display in the header of the dialog page "MONITORING"
switches to
* BUSY *
, the start time of the measurement is entered un-
der
Determination
and the determination is given a standard name,
also formed from the start time. Further, during the measurement the
measurement window is superimposed which shows the current pro-
gram line being executed and the remaining line time.
•
Shortly after the start of the determination, the screen shows the prompt
to add 10 mL sample to the measuring vessel.
5
Addition of sample
•
Remove stopper 25
25 on 747 VA Stand and use a pipette to add 10 mL of
drinking water to the measuring vessel through the pipetting opening.
Press the
•
The screen shows the prompt to add 1 mL acetate buffer to the measur-
ing vessel.
6
Addition of acetate buffer
•
Use a pipette to add 1 mL acetate buffer (1 mol/L, pH 4.6) to the
measuring vessel through the pipetting opening. Then close the pipet-
ting opening using stopper 25
25 and press the
•
The screen shows the prompt to add 100
µ
L of the solution
CdRefSol
with the internal standard to the measuring vessel.
7
Addition of the internal standard
•
Use a pipette to add 100
µµL of the cadmium standard solution with a
mass concentration of 1 mg/L to the measuring vessel through the pi-
petting opening. Then close the pipetting opening using stopper 25
25 and
press the
8
Measurement of the sample solution
•
Measurement of the sample solution runs as follows (see also section
4.4.3):
−
Purging and stirring for 180 s.
−
Call-up of the segment
DPASV
for measured value recording.
Lead and cadmium are enriched with stirring at the hanging mer-
cury drop at a measurement voltage of –800 mV and then deter-
mined anodically without stirring from –800 ... –200 mV. The se-
lected measurement mode DP leads to a current peak for lead
whose maximum lies at ca. –400 mV and a current peak for cad-
mium with a maximum at ca. –580 mV.
−
The
DPASV
segment is then repeated twice. Both the three single
measurements for the substance lead and also the three single
measurements for cadmium are allocated the VR numbers
00
,
01
and
02
.
•
On completion of the last measurement, the lead mass concentration is
calculated using the proportionality factor stored in the method. The pro-
gram is now at an end. The status display in the header of the dialog
page "MONITORING" switches back to
* READY *
, the instrument is
ready for a new determination.