7 7.3.2 list of error messages – Metrohm 746 VA Trace Analyzer User Manual
Page 506

7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-7
7.3.2
List of error messages
The following list contains all error messages in alphabetical order. You will find ex-
planations regarding how to proceed should the instrument need to be initialized
(reset or overall initialization) in section 7.3.3. The abbreviations in the ”Exit” column
have the following meaning:
Exit
Meaning with ”
Auto.batch = off
”
Meaning with ”
Auto.batch = on
”
E
Exit with
no program has been started.
–
E/C
Exit with
a program already started continues.
The error message is displayed for
10 s and is then acknowledged
automatically with
started program continues to run.
E/S
Exit with
a program already started is stopped.
The error message is displayed for
10 s and is then acknowledged
automatically with
started program is stopped.
E/N
Exit with
a program already started is stopped.
The error message is displayed for
10 s and is then acknowledged
automatically with
started program is stopped. The next
determination is then started.
S
The instrument is blocked, exit is no
longer possible; the instrument must
be reinitialized (see section 7.3.3).
The instrument is blocked, exit is no
longer possible; the instrument must
be reinitialized (see section 7.3.3).
r/a
Exit with
(a = abort).
The error message is displayed for
10 s and is then acknowledged
automatically with
Esc
Exit with
for the acknowledgment of error
messages in the error window.
–