5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-196

A>M

Page:

O, S

Stand:

A

Abbreviation:

A>

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. A>M ..... addition AUX -> MEAS V.add ........mL

Function

Transfer solution from auxiliary vessel (AUX) on the 695 Autosampler to
the measuring vessel (MEAS) on the 747 VA Stand.

Parameter

Range; Default value / Meaning

V.add

0.010...50.000 mL; 0.100 mL

Transfer volume

Solution volume that should be transferred from the auxiliary vessel on
the sample rack of the 695 Autosampler to the measuring vessel on the
747 VA Stand.

Description

The

A>M

instruction is used to transfer the solution volume specified

under

V.add

from the auxiliary vessel (AUX) on the 695 Autosampler to

the measuring vessel (MEAS) on the 747 VA Stand.

Comments

The

A>M

instruction can be used only in operation with the 695

Autosampler.

If the solution volume in the auxiliary vessel <

V.add

, an error

message appears.

see also

F>A

,

F>M