5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-236

Description

The

EXIT

instruction can be inserted in the following loops or blocks:

(ADD ... ADD)n
(CAL ... CAL)n
(REP ... REP)n
(VAR ... VAR)n
(CASE ... CASE)

On the basis of test condition contained in the

EXIT

instruction as a

parameter (see

(Test)

), a decision is made during the program run

whether or not the loop or block should be exited:

If the test condition is met (true), a jump is made to the

end of the loop or block.

If the test condition is not met (false), the instructions

which follow in the loop or block are executed.

Comments

In nested loops or blocks a jump is always made to the end of that
loop or block in which the

EXIT

instruction is located. If an exit from

all loops or blocks is desired, the

EXIT

instruction must be entered

for every loop or every block.

For test conditions with evaluation values

EV:

or mass

concentrations

MC:

, it must be borne in mind that these

values are set to 0 if no evaluation was possible. This can be used for
detection of this specific case (see example).

Example

see

(CASE

see also

(CASE, CASE), CASE, (ADD, (CAL, (REP, (VAR