5.3 Set Up

746 VA Trace Analyzer / 747 VA Stand

5-26

Parameter

Range; Default value / Meaning

V.meas max

1...999 mL; 80 mL

Maximum fill volume

Setting of the maximum fill volume for the measuring vessel of the 747
VA Stand. It must be ensured here that the measuring vessel is not
overfilled. In dispensing operations which would lead to this volume
being exceeded, an error message appears.

The following settings are recommended:

Measuring vessel

V.meas max

Measuring vessel

V.meas max

6.1415.210

80 mL

6.1450.210

80 mL

6.1415.150

70 mL

6.1456.210

80 mL

6.1418.220

70 mL

6.1457.210

80 mL

V.rinse

1...999 mL; 8 mL

Rinse volume

Setting of the rinse volume for manual rinsing of the measuring vessel
triggered by the softkey

.

This parameter is displayed only if two rinsing Dosinos are attached to
the 746 VA Trace Analyzer.

V.standby

1...999 mL; 0 mL

Standby volume

Setting of the volume of rinsing solution which is filled into the meas-
uring vessel after manual rinsing triggered by the softkey

meas.v.>

. This volume can be used as conditioning solution for the

electrodes.

This parameter is displayed only if two rinsing Dosinos are attached to
the 746 VA Trace Analyzer.

U.standby

display only

Standby voltage at electrodes

Display of the polarization voltage set to the electrodes at the end of
the determination (

U.standby

is set with the

ØMEAS

command).

Rinse unit pos.

exch., rinse; exch.

Position of the rinsing Dosinos

Display of the current position of the two rinsing Dosinos used for
rinsing the measuring vessel and siphoning off solution:

exch.

Exchange position (dosing unit can be changed)

rinse

Rinse position

To fill or empty the containers or to remove the dosing units from the
Dosinos, the Dosinos must be in the

exch.

position. If the Dosinos

are in the

rinse

position, they have to be switched to the exchange

position by pressing

.

This parameter is displayed only if two rinsing Dosinos are attached to
the 746 VA Trace Analyzer.