237 f>a, O, s – Metrohm 746 VA Trace Analyzer User Manual
Page 388

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-237
F>A
Page:
O, S
Stand:
A
Abbreviation:
F
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. F>A ..... addition FEED -> AUX V.add ........mL
Function
Transfer solution from sample vessel (FEED) to the auxiliary vessel
(MEAS) on the 695 Autosampler.
Parameter
Range; Default value / Meaning
V.add
0.010...50.000 mL; 1.000 mL
Transfer volume
Solution volume that should be transferred from the sample vessel on
the sample rack of the 695 Autosampler to the auxiliary vessel.
Description
The
F>A
instruction transfers the solution volume specified under
V.add
from the sample vessel (FEED) to the auxiliary vessel (AUX) on the 695
Autosampler.
Comments
•
The
F>A
instruction can be used only in operation with the 695
Autosampler.
•
The auxiliary vessel (AUX) must be located on the sample rack
immediately after the sample vessel (FEED).
•
If the solution volume in the sample vessel <
V.add
, an error
message appears.
see also
A>M
,
F>M
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)