5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-237

F>A

Page:

O, S

Stand:

A

Abbreviation:

F

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. F>A ..... addition FEED -> AUX V.add ........mL

Function

Transfer solution from sample vessel (FEED) to the auxiliary vessel
(MEAS) on the 695 Autosampler.

Parameter

Range; Default value / Meaning

V.add

0.010...50.000 mL; 1.000 mL

Transfer volume

Solution volume that should be transferred from the sample vessel on
the sample rack of the 695 Autosampler to the auxiliary vessel.

Description

The

F>A

instruction transfers the solution volume specified under

V.add

from the sample vessel (FEED) to the auxiliary vessel (AUX) on the 695
Autosampler.

Comments

The

F>A

instruction can be used only in operation with the 695

Autosampler.

The auxiliary vessel (AUX) must be located on the sample rack
immediately after the sample vessel (FEED).

If the solution volume in the sample vessel <

V.add

, an error

message appears.

see also

A>M

,

F>M