beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 537

background image

7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-38

Standard addition curves are not reproducible or non-linear

Check program in the instrument (stirring time, etc.).

Check MME, change capillary if necessary (see section 3.4.8) or replace

sealing needle (see section 3.4.9).

The pipetting process was not correct: was the pipetting unit used properly?

Repeat the analysis again or try out automatic standard addition with Dosi-

mat, Dosino or 6.5611.010 Pipetting equipment.

Pipetting the standard solutions must be carried out by one and the same

person or with the same instrument or the same pipette.

Check and test pipetting process. When were the pipettes last calibrated

(GLP)?

Organic components interfere with the analysis: carry out a UV digestion or

other suitable sample preparation.

Are the calibration solutions too old?

Would a calibration curve be more suitable?

The linearity at the HMDE is naturally not as good as with the DME. The linear

range is in general no larger than 1 - 2 decades.

Peak displacement

Check and adjust the pH of the solution.

Check electrolyte composition and correct if necessary.

Organic components interfere with the analysis: carry out a UV digestion or

other suitable sample preparation.

Carry out a standard addition to check if the correct peak is present.

Enter a new half-wave potential in the instrument and recalculate the results.

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

No peak found

The peak is only displaced: adjust the half-wave potential and recalculate.

The sample concentration is too small: increase the sample volume.

Has the complex former been forgotten? (adsorptive voltammetry).

The deposition time under

Meas

in "OP.SEQUENCE" in the inverse voltam-

metry is too short: increase the time.

No Hg drops at the capillary: check MME. Adjust sealing needle (see section

3.4.4/3.4.5). If necessary, change capillary (see section 3.4.8) or replace

sealing needle (see section 3.4.9).

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

Are organic components present? Carry out a UV digestion or other suitable

sample preparation.