5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-244

HOLD

Page:

O, S

Stand:

M, A

Abbreviation:

HO

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. HOLD ..... Message ................

Function

Program hold with output of a message.

Parameter

Range; Default value / Meaning

Message

16 characters; -

Message

The message put in here is displayed in the screen window

MEAS.TASK

MESSAGE

when the

HOLD

instruction is reached. The program is not

continued until this message has been confirmed with the
key.

Description

The

HOLD

instruction is used to interrupt the program at a specific

location. The program is not continued until the message

Message

displayed on the screen associated with

HOLD

has been confirmed with

the key.