4.3 Lead determination with calibration curve

746 VA Trace Analyzer / 747 VA Stand

4-33

Method: CALCRV SUBSTANCES

Pb - DPASV

Calibration 1996-11-22 16:35 Coefficients

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДД

Technique cal.crv. Y.offset -8.183e-10

Curve type linear Slope 0.002712

Temperature 20.0 °C Nonlin

Mean dev. 1.113e-09

Range min g/L # CSoln.name Mass conc.

Range max g/L ДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДД

0 Pb-1 2.5 ug/L

M.conc./cm g/L 1 Pb-2 5 ug/L

M.conc.begin zero 2 Pb-3 10 ug/L

3 Pb-4 20 ug/L

V0.CSoln. 10.000 mL 4 g/L

5 g/L

6 g/L

7 g/L

8 g/L

9 g/L

Modify

column

Modify

line

Select

subst.

Link to

segment

Delete

subst.

Subst.

name

>>

The following is a brief explanation of the most important parameters:

Technique

Calibration technique: calibration curve

Curve type

Curve type for calculation of the calibration curve:
linear

Temperature

Temperature of the analysis solution

V0.CSoln.

Volume added of every calibration solution

Y.offset

Offset of the calibration curve

Slope

Slope of the calibration curve

Mean dev.

Mean deviation of the calibration curve

CSoln.name

Name of the calibration solutions

Mass conc.

Mass concentration of the calibration solutions

4

Switch to dialog page "CALCULATION"

Press the yellow key. The dialog page "CALCU-

LATION" appears for calculation of the final results. The mass concen-

tration of lead (

MC:Pb

) determined using the calibration curve is put out

as the result in g/L with 4 significant figures.

5

Switch to dialog page "DOCUMENTATION"

Press the yellow key. The dialog page "DOCU-

MENTATION" appears for the automatic output of reports and curves.

The two lines entered in the example have the following meaning:

Report Full

Printout of the full result report on the printer

Curve Cal Sbst:Pb

Printout of the calibration curve on the printer

Press the softkey

. The subpage of "DOCUMENTATION"

appears with the method comment describing, among other things, the

procedure used in the calibration and determination.

Finally press the key to return to the main page of

"DOCUMENTATION".